Revised 8/27/13
3-17
Performing a Cortical Stimulation Study
Start the study
NOTE:
You can periodically verify the stimulator is operating properly by
pressing the
Check Stim
button. This delivers a single bi-phasic pulse to a load
and displays the requested current Stimulus Level.
1. When ready to begin the study, verify that the stimulus parameters are
correct on the Cortical Stimulus panel and then press the Stimulus Control
Unit (SCU)
Start
button to deliver the stimulation to the patient.
The following information is displayed on the Stimulus Control Unit (SCU)
display:
•
Level of stimulation
being delivered to the patient is displayed. If the
level is 20% or more higher than the selected level, stimulation is stopped
automatically and an error report is generated.
2. If you wish during surgery, when using the Bipolar Probe you can select the
Continuous mode by pressing the
Train
Duration
button and selecting
Continuous
using the
Selector
knob.
NOTE:
If you change any parameters on the Stimulus Control Unit (SCU),
stimulation will stop being delivered to the patient. When done making
adjustments and
Ready to stim
is displayed to the right of
Status
on the Stimulus
Control Unit (SCU), press the
Start
button again to resume stimulation.
Summary of Contents for Nicolet SCU
Page 2: ......
Page 9: ...Revised 8 27 13 1 1 1 General Information...
Page 10: ...Nicolet Cortical Stimulator 1 2 Revised 8 27 13 Blank Page...
Page 21: ...Revised 8 27 13 2 1 2 Nicolet Cortical Stimulator Overview...
Page 22: ...Nicolet Cortical Stimulator 2 2 Revised 8 27 13 Blank Page...
Page 37: ...Revised 8 27 13 3 1 3 Performing a Cortical Stimulation Study...
Page 38: ...Nicolet Cortical Stimulator 3 2 Revised 8 27 13 Blank Page...
Page 55: ...Revised 8 27 13 4 1 4 Electromagnetic Compatibility EMC...
Page 56: ...Nicolet Cortical Stimulator 4 2 Revised 8 27 13 Blank Page...
Page 62: ...Nicolet Cortical Stimulator 4 8 Revised 8 27 13 Blank page...
Page 63: ...Revised 8 27 13 5 1 5 Error Codes...
Page 64: ...Nicolet Cortical Stimulator 5 2 Revised 8 27 13 Blank Page...
Page 66: ...Nicolet Cortical Stimulator 5 4 Revised 8 27 13 Blank page...