Operating Environment
Caution
The operating temperature must not be exceeded, even when used in a
chassis with a higher temperature range.
Ambient temperature range
5 °C to 50 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 high temperature
limit.)
Extended temperature option
0 °C to 55 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 low temperature
limit and MIL-PRF-28800F Class 2 high
temperature limit.)
Relative humidity range
10% to 90%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
Storage Environment
Ambient temperature range
-40 °C to 65 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 low temperature
limit.)
Extended Temperature Option
-40 °C to 71 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3 limits.)
Relative humidity range
5% to 95%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
Shock and Vibration
Operating shock
30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC-60068-2-27. Meets MIL-
PRF-28800F Class 2 limits.)
Random vibration
Operating
5 Hz to 500 Hz, 0.3 g
rms
(with solid-state hard
drive)
Nonoperating
5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC-60068-2-64. Nonoperating test
profile exceeds the requirements of MIL-
PRF-28800F, Class 3.)
NI PXIe-8840 Quad-Core User Manual
|
© National Instruments
|
25