![background image](http://html.mh-extra.com/html/national-instruments/pxie-5113/pxie-5113_calibration-procedure_710211021.webp)
Table 5. 50
Ω
Bandwidth Verification (Continued)
Config
Sample
Rate
(S/s)
Vertical
Range
(V
pk-pk
)
Maximum
Input
Frequency
(MHz)
Test Point
As-
Found
Limits
(dB)
As-Left
Limits
(dB)
Frequency
(MHz)
Amplitude
(dBm)
47
3 G
2 V
500
0.05
7.0
REF
REF
48
3 G
2 V
500
475.1
7.0
-3.0 to
1.3
-2.2 to
1.3
49
1.5 G
4 V
350
0.05
9.0
REF
REF
50
1.5 G
4 V
350
325.1
9.0
-3.0 to
1.3
-2.6 to
1.3
51
3 G
4 V
350
0.05
9.0
REF
REF
52
3 G
4 V
350
325.1
9.0
-3.0 to
1.3
-2.6 to
1.3
53
1.5 G
4 V
500
0.05
9.0
REF
REF
54
1.5 G
4 V
500
475.1
9.0
-3.0 to
1.3
-2.2 to
1.3
55
3 G
4 V
500
0.05
9.0
REF
REF
56
3 G
4 V
500
475.1
9.0
-3.0 to
1.3
-2.2 to
1.3
57
1.5 G
10 V
350
0.05
9.0
REF
REF
58
1.5 G
10 V
350
325.1
9.0
-3.0 to
1.3
-2.6 to
1.3
59
3 G
10 V
350
0.05
9.0
REF
REF
60
3 G
10 V
350
325.1
9.0
-3.0 to
1.3
-2.6 to
1.3
61
1.5 G
10 V
500
0.05
9.0
REF
REF
62
1.5 G
10 V
500
475.1
9.0
-3.0 to
1.3
-2.3 to
1.3
PXIe-5113 Calibration Procedure
|
© National Instruments
|
21