4.
Connect an SMA (m)-to-BNC (m) adapter to one of the power splitter output ports. Refer
to this assembly as
splitter output 1
.
5.
Connect the 50 Ω BNC terminator (f) to splitter output 1.
6.
Connect the other SMA (m)-to-BNC (m) adapter to the other output port of the power
splitter. Refer to this assembly as
splitter output 2
.
7.
Connect the power sensor to splitter output 2.
The following figure illustrates the hardware setup.
1
6
4
5
3
7
8
2
5
1. Signal generator
2. SMA (f)-to-N (m) adapter
3. SMA (m)-to-SMA (m) cable
4. 50
Ω
BNC terminator (f)
5. SMA (m)-to-BNC (m) adapter
6. Power splitter
7. BNC (f)-to-N (f) adapter
8. Power sensor
8.
Configure the power sensor with the following settings:
•
Power measurement: continuous average
•
Path selection: automatic
•
Averaging: automatic
•
Averaging resolution: 4 (0.001 dB)
•
Aperture: 20 ms