Table 4. DC Accuracy Verification (Continued)
Config
Input
Impedance
(
Ω
)
Vertical
Range
(V
pk-pk
)
Vertical
Offset
(V)
Sample
Rate
(GS/s)
Test
Points
(V)
As-
Found
Limits
(mV)
As-Left
Limits (mV)
99
1 M
4 V
0
1.5
-1.9
±78.2
±46.7
100
1 M
4 V
0
3.0
-1.9
±78.2
±46.7
101
1 M
4 V
20
1.5
21.9
±158.2
±82.7
102
1 M
4 V
20
3.0
21.9
±158.2
±82.7
103
1 M
4 V
-20
1.5
-21.9
±158.2
±82.7
104
1 M
4 V
-20
3.0
-21.9
±158.2
±82.7
105
1 M
10 V
0
1.5
4.75
±195.2
±116.75
106
1 M
10 V
0
3.0
4.75
±195.2
±116.75
107
1 M
10 V
0
1.5
-4.75
±195.2
±116.75
108
1 M
10 V
0
3.0
-4.75
±195.2
±116.75
109
1 M
10 V
100
1.5
104.75
±595.2
±296.75
110
1 M
10 V
100
3.0
104.75
±595.2
±296.75
111
1 M
10 V
-100
1.5
-104.75
±595.2
±296.75
112
1 M
10 V
-100
3.0
-104.75
±595.2
±296.75
113
1 M
20 V
0
1.5
9.5
±390.2
±233.5
114
1 M
20 V
0
3.0
9.5
±390.2
±233.5
115
1 M
20 V
0
1.5
-9.5
±390.2
±233.5
116
1 M
20 V
0
3.0
-9.5
±390.2
±233.5
117
1 M
20 V
100
1.5
109.5
±790.2
±413.5
118
1 M
20 V
100
3.0
109.5
±790.2
±413.5
119
1 M
20 V
-100
1.5
-109.5
±790.2
±413.5
120
1 M
20 V
-100
3.0
-109.5
±790.2
±413.5
121
1 M
40 V
0
1.5
19
±780.2
±467
16
|
ni.com
|
PXIe-5113 Calibration Procedure