Verifying DC Accuracy
This procedure verifies the DC accuracy of the PXIe-5113 by comparing the voltage measured
by the device to the value sourced by the voltage standard.
Caution
Avoid touching the connections when generating a high voltage from the
calibrator.
Refer to the following table as you complete the following steps.
Table 4. DC Accuracy Verification
Config
Input
Impedance
(
Ω
)
Vertical
Range
(V
pk-pk
)
Vertical
Offset
(V)
Sample
Rate
(GS/s)
Test
Points
(V)
As-
Found
Limits
(mV)
As-Left
Limits (mV)
1
50
0.1 V
0
1.5
0.0475
±2.15
±1.098
2
50
0.1 V
0
3.0
0.0475
±2.15
±1.10275
3
50
0.1 V
0
1.5
-0.0475
±2.15
±1.098
4
50
0.1 V
0
3.0
-0.0475
±2.15
±1.10275
5
50
0.1 V
4.95
1.5
4.9975
±21.95
±10.008
6
50
0.1 V
4.95
3.0
4.9975
±21.95
±10.01275
PXIe-5113 Calibration Procedure
|