Table 5. 50
Ω
Bandwidth Verification (Continued)
Config
Sample
Rate
(S/s)
Vertical
Range
(V
pk-pk
)
Maximum
Input
Frequency
(MHz)
Test Point
As-
Found
Limits
(dB)
As-Left
Limits
(dB)
Frequency
(MHz)
Amplitude
(dBm)
63
3 G
10 V
500
0.05
9.0
REF
REF
64
3 G
10 V
500
475.1
9.0
-3.0 to
1.3
-2.3 to
1.3
1.
Connect splitter output 2 of the power sensor assembly from the
section to channel 0 of the PXIe-5113.
Note
The power sensor assembly must match the configuration used in the
section, in which the power sensor is connected to
splitter output 1 and the signal generator is connected to the input port of the
power splitter.
The following figure illustrates the hardware setup.
22
|
ni.com
|
PXIe-5113 Calibration Procedure