background image

©

 National Instruments

9

NI PXIe-4353 Calibration Procedure

8.

Configure the timing properties for the voltage acquisition. Refer to Table 3.

9.

Start the acquisition.

10. Acquire the voltage data.

Note

Make sure that the calibrator and the NI PXIe-4353 are completely settled on the desired 

resistance before acquiring data. When using the Fluke 5520A, wait at least 5000 ms for the system 
to settle before issuing the start task function to the NI PXIe-4353.

11. Calculate the equivalent resistance using the following equation:

where 

V

CJC_ave

 is the average of the voltage values for the channel you acquired.

12. Compare the resulting average to the 

Upper Limit

 and 

Lower Limit

 values in Table 7. If the result 

is between these values, the channel passes the test.

13. Clear the acquisition.

14. Repeat steps through 12 for all CJC channels and test point values in Table 7. NI recommends 

that you verify all values, although you can save time by verifying only the values used in your 
application.

15. Close the DAQmx calibration session with the action Cancel.

16. Disable the calibrator output and disconnect the calibrator from the CAL-4353.

Note

After verifying your module, you can update the last calibration date by opening a DAQmx 

external calibration session and closing it with the action Commit.

Table 3.  

DAQmx Timing Function Rate and Samples per Channel for CJC Verification

Timing Mode

Rate (S/s)

Samples per Channel

1

17

10

2

34

10

3

68

20

4

90

35

5

90

50

6

90

80

7

90

150

R

CJC

40 k

Ω

2.048 V

V

CJC_ave

-------------------

1

----------------------------

=

Summary of Contents for PXIe-4353

Page 1: ...ts 13 Module Verification Test Limit Tables 13 Where to Go for Support 15 Software Install NI DAQmx 9 3 or later on the calibration computer NI DAQmx includes high level function calls to simplify the...

Page 2: ...u use the instruments in Table 1 for calibrating an NI PXIe 4353 module Table 1 Recommended Equipment Equipment Recommended Model Requirements Voltage calibrator Fluke 5520A If this instrument is unav...

Page 3: ...re This section provides instructions for verifying the voltage accuracy performance of the NI PXIe 4353 The calibration process consists of the following steps 1 Initial Setup for the NI PXIe 4353 In...

Page 4: ...ration adjustment can cause you to incorrectly calibrate your module The onboard temperature sensor on the module can be used to verify that this condition does not exist Complete the following steps...

Page 5: ...4353 while referring to Figure 1 Figure 1 CAL 4353 to Fluke 5520A Calibrator Connections 2 Open a DAQmx calibration session on your device 3 Call the DAQmx Connect SC Express Calibration Accessory Cha...

Page 6: ...els you want to verify 7 Set the autozero mode to Every Sample and disable open thermocouple detection 8 On the DAQmx Channel property node set the DAQmx ADC timing mode property to Custom and set the...

Page 7: ...e of the analog input offset of the module 1 Open a DAQmx calibration session on your device 2 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the...

Page 8: ...bration session on your device 4 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the connection to _cjc_calibration and the physical channels to t...

Page 9: ...Limit and Lower Limit values in Table 7 If the result is between these values the channel passes the test 13 Clear the acquisition 14 Repeat steps 5 through 12 for all CJC channels and test point valu...

Page 10: ...mx Connect SC Express Calibration Accessory Channels VI Set the connection to _tc_calibration and the physical channels to Dev1 ai0 31 4 Set the calibrator to output 70 mV and wait for the calibrator...

Page 11: ...0 Set the resistance calibrator to 2 wire compensation mode 11 Configure the CAL 4353 using the DAQmx Connect SC Express Calibration Accessory Channels VI Set the connection to _cjc_calibration and th...

Page 12: ...ose External Calibration VI and set the action to Commit This VI also saves the date time and temperature of the adjustment to the onboard memory Note If an error occurs during adjustment no constants...

Page 13: ...its Contains the Upper Limits and Lower Limits for the test point value That is when the module is within its 1 year calibration interval the test point value should fall between these upper and lower...

Page 14: ...02 3 0 000002 0 000002 4 0 000003 0 000003 5 0 000005 0 000005 6 0 0000065 0 0000065 7 High Speed 0 0000075 0 0000075 Table 6 NI PXIe 4353 Offset with Autozero Disabled Verification Test Limits Timing...

Page 15: ...side the United States visit the Worldwide Offices section of ni com niglobal to access the branch office Web sites which provide up to date contact information support phone numbers email addresses a...

Page 16: ...countries Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the approp...

Reviews: