background image

NI PXIe-4353 Calibration Procedure

2

ni.com

Documentation

The following documents are your primary references for writing your calibration utility with 
NI-DAQmx. You can download the latest version of these documents from the NI Web site at 

ni.com/manuals

.

The 

NI SC Express 4353 Installation Guide and Terminal Block Specifications

 provides 

instructions for installing and configuring the NI PXIe-4353 module and TB-4353 accessory.

The 

NI PXIe-4353 User Manual

 describes how to use the NI PXIe-4353.

The 

NI PXIe-4353 Specifications

 lists the specifications for the NI PXIe-4353.

The 

CAL-4353 Specifications

 lists the specifications for the CAL-4353.

The 

NI-DAQmx Help

 includes information about creating applications that use the NI-DAQmx 

driver.

The 

NI-DAQmx C Reference Help

 includes information about the functions in the driver.

Calibration Interval

National Instruments recommends a calibration interval of one year for the NI PXIe-4353. You should 
adjust the recommended calibration interval based on the measurement accuracy demands of your 
application.

Test Equipment

National Instruments recommends that you use the instruments in Table 1 for calibrating an 
NI PXIe-4353 module.

Table 1.  

Recommended Equipment

Equipment

Recommended Model

Requirements

Voltage calibrator

Fluke 5520A

If this instrument is unavailable for voltage accuracy verification and 
adjustment, use a high-precision voltage source with an accuracy of at 
least 70 ppm when sourcing up to 50 

μ

A.

Resistance calibrator

*

Fluke 5520A

If this instrument is unavailable for resistance accuracy verification and 
adjustment, use a high-precision resistance source with an accuracy of 
at least 150 ppm 2-wire compensation.

If the resistance source does not have 2-wire compensation, the lead 
wire resistance must be included in the total resistance source accuracy.

PXI Express chassis

NI PXIe-1062Q

Calibration accessory

CAL-4353

This accessory is used for module verification and adjustment.

Connecting wires

Banana-jack 
terminated copper wire

Copper wire is used for validating the voltage accuracy and adjustment 
of the analog input channels. Copper wire is also used for validating the 
resistance accuracy and calibration of the CJC resistance input 
channels.

When connecting the resistance calibrator, keep the lead wires 1.0

Ω

 less than the maximum lead compensation specification of the resistance 

calibrator. When using the Fluke 5520A, keep the lead wires <1.1

Ω

.

Summary of Contents for PXIe-4353

Page 1: ...ts 13 Module Verification Test Limit Tables 13 Where to Go for Support 15 Software Install NI DAQmx 9 3 or later on the calibration computer NI DAQmx includes high level function calls to simplify the...

Page 2: ...u use the instruments in Table 1 for calibrating an NI PXIe 4353 module Table 1 Recommended Equipment Equipment Recommended Model Requirements Voltage calibrator Fluke 5520A If this instrument is unav...

Page 3: ...re This section provides instructions for verifying the voltage accuracy performance of the NI PXIe 4353 The calibration process consists of the following steps 1 Initial Setup for the NI PXIe 4353 In...

Page 4: ...ration adjustment can cause you to incorrectly calibrate your module The onboard temperature sensor on the module can be used to verify that this condition does not exist Complete the following steps...

Page 5: ...4353 while referring to Figure 1 Figure 1 CAL 4353 to Fluke 5520A Calibrator Connections 2 Open a DAQmx calibration session on your device 3 Call the DAQmx Connect SC Express Calibration Accessory Cha...

Page 6: ...els you want to verify 7 Set the autozero mode to Every Sample and disable open thermocouple detection 8 On the DAQmx Channel property node set the DAQmx ADC timing mode property to Custom and set the...

Page 7: ...e of the analog input offset of the module 1 Open a DAQmx calibration session on your device 2 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the...

Page 8: ...bration session on your device 4 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the connection to _cjc_calibration and the physical channels to t...

Page 9: ...Limit and Lower Limit values in Table 7 If the result is between these values the channel passes the test 13 Clear the acquisition 14 Repeat steps 5 through 12 for all CJC channels and test point valu...

Page 10: ...mx Connect SC Express Calibration Accessory Channels VI Set the connection to _tc_calibration and the physical channels to Dev1 ai0 31 4 Set the calibrator to output 70 mV and wait for the calibrator...

Page 11: ...0 Set the resistance calibrator to 2 wire compensation mode 11 Configure the CAL 4353 using the DAQmx Connect SC Express Calibration Accessory Channels VI Set the connection to _cjc_calibration and th...

Page 12: ...ose External Calibration VI and set the action to Commit This VI also saves the date time and temperature of the adjustment to the onboard memory Note If an error occurs during adjustment no constants...

Page 13: ...its Contains the Upper Limits and Lower Limits for the test point value That is when the module is within its 1 year calibration interval the test point value should fall between these upper and lower...

Page 14: ...02 3 0 000002 0 000002 4 0 000003 0 000003 5 0 000005 0 000005 6 0 0000065 0 0000065 7 High Speed 0 0000075 0 0000075 Table 6 NI PXIe 4353 Offset with Autozero Disabled Verification Test Limits Timing...

Page 15: ...side the United States visit the Worldwide Offices section of ni com niglobal to access the branch office Web sites which provide up to date contact information support phone numbers email addresses a...

Page 16: ...countries Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the approp...

Reviews: