background image

NI PXIe-4353 Calibration Procedure

4

ni.com

Initial Setup for the NI PXIe-4353

You must configure the module in Measurement & Automation Explorer (MAX) to communicate with 
NI-DAQmx.

Complete the following steps to configure a module in MAX. Refer to the 

NI SC Express 4353 

Installation Guide and Terminal Block Specifications

 for complete installation instructions.

1.

Install the NI-DAQmx driver software.

Caution

Always have the PXI Express chassis powered off when inserting a module.

2.

Insert the module into an available slot in the PXI Express chassis.

3.

Attach the CAL-4353 to the NI PXIe-4353.

Note

The CAL-4353 must be within its calibration interval and if it is not, you must calibrate the 

CAL-4353 before calibrating the NI PXIe-4353 module.

4.

Power on the chassis.

5.

Launch MAX.

6.

Expand 

Devices and Interfaces

 to confirm that MAX detects the module and terminal block. 

The terminal block should appear beneath its associated module.

7.

Right-click 

NI PXIe-4353

 and select 

Self-Test

 to ensure that the module is working properly.

Note

You can also perform the self-test operation programmatically, using the DAQmx Self-Test 

Device VI or the equivalent C/.NET functionality.

Verifying Module Operating Temperature Changes

A module operating temperature greater than ±10 °C since the previous calibration adjustment can 
cause you to incorrectly calibrate your module. The onboard temperature sensor on the module can be 
used to verify that this condition does not exist. Complete the following steps to compare the current 
module temperature to the temperature measured during the last calibration.

1.

Launch MAX.

2.

Select 

My System»Devices and Interfaces»Chassis 1»NI PXIe-4353

.

Note

MAX assigns a chassis name for each chassis in an application. This document uses Chassis 1. 

In your application, use the appropriate name assigned to your chassis by MAX.

3.

Click the 

Calibration

 tab.

If the difference between the current operating temperature and the operating temperature from the 
last calibration adjustment is greater than 10 °C, the limits in the calibration tables are not valid.

Note

You can also query the temperatures programmatically, using the DAQmx Calibration Info 

property node or the equivalent C/.NET functionality.

Note

The temperature reading returned by the module is an internal temperature. Typically, the 

internal temperature is approximately 10 °C warmer than the ambient temperature.

If the module temperature difference is outside the maximum range, you should choose one of the 
following options:

Recalculate the test limits to include the additional error due to temperature drift. Refer to the 

NI PXIe-4353 Specifications 

for more information.

Change the system so that the operating temperature will be closer to the temperature recorded 
during the last calibration adjustment.

Summary of Contents for PXIe-4353

Page 1: ...ts 13 Module Verification Test Limit Tables 13 Where to Go for Support 15 Software Install NI DAQmx 9 3 or later on the calibration computer NI DAQmx includes high level function calls to simplify the...

Page 2: ...u use the instruments in Table 1 for calibrating an NI PXIe 4353 module Table 1 Recommended Equipment Equipment Recommended Model Requirements Voltage calibrator Fluke 5520A If this instrument is unav...

Page 3: ...re This section provides instructions for verifying the voltage accuracy performance of the NI PXIe 4353 The calibration process consists of the following steps 1 Initial Setup for the NI PXIe 4353 In...

Page 4: ...ration adjustment can cause you to incorrectly calibrate your module The onboard temperature sensor on the module can be used to verify that this condition does not exist Complete the following steps...

Page 5: ...4353 while referring to Figure 1 Figure 1 CAL 4353 to Fluke 5520A Calibrator Connections 2 Open a DAQmx calibration session on your device 3 Call the DAQmx Connect SC Express Calibration Accessory Cha...

Page 6: ...els you want to verify 7 Set the autozero mode to Every Sample and disable open thermocouple detection 8 On the DAQmx Channel property node set the DAQmx ADC timing mode property to Custom and set the...

Page 7: ...e of the analog input offset of the module 1 Open a DAQmx calibration session on your device 2 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the...

Page 8: ...bration session on your device 4 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the connection to _cjc_calibration and the physical channels to t...

Page 9: ...Limit and Lower Limit values in Table 7 If the result is between these values the channel passes the test 13 Clear the acquisition 14 Repeat steps 5 through 12 for all CJC channels and test point valu...

Page 10: ...mx Connect SC Express Calibration Accessory Channels VI Set the connection to _tc_calibration and the physical channels to Dev1 ai0 31 4 Set the calibrator to output 70 mV and wait for the calibrator...

Page 11: ...0 Set the resistance calibrator to 2 wire compensation mode 11 Configure the CAL 4353 using the DAQmx Connect SC Express Calibration Accessory Channels VI Set the connection to _cjc_calibration and th...

Page 12: ...ose External Calibration VI and set the action to Commit This VI also saves the date time and temperature of the adjustment to the onboard memory Note If an error occurs during adjustment no constants...

Page 13: ...its Contains the Upper Limits and Lower Limits for the test point value That is when the module is within its 1 year calibration interval the test point value should fall between these upper and lower...

Page 14: ...02 3 0 000002 0 000002 4 0 000003 0 000003 5 0 000005 0 000005 6 0 0000065 0 0000065 7 High Speed 0 0000075 0 0000075 Table 6 NI PXIe 4353 Offset with Autozero Disabled Verification Test Limits Timing...

Page 15: ...side the United States visit the Worldwide Offices section of ni com niglobal to access the branch office Web sites which provide up to date contact information support phone numbers email addresses a...

Page 16: ...countries Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the approp...

Reviews: