background image

©

 National Instruments

3

NI PXIe-4353 Calibration Procedure

Test Conditions

The following setup and environmental conditions are required to ensure the module meets published 
specifications.

Keep connections to the module as short as possible. Long cables and wires act as antennas, picking 
up extra noise that can affect measurements.

Use shielded wire for all cable connections to the module. Use twisted-pair wire to eliminate noise.

Maintain an ambient temperature of 23 °C ±5 °C. The module temperature will be greater than the 
ambient temperature.

Keep relative humidity below 80%.

Allow a warm-up time of 15 minutes to ensure that the measurement circuitry of the NI PXIe-4353 
and CAL-4353 are at a stable operating temperature.

Note

If you are using the Fluke 5520A, allow the calibrator to warm up for 30 minutes and maintain 

an ambient temperature of 23 °C ±3 °C.

1

Ensure that the PXI Express chassis fan speed is set to HIGH, that the fan filters are clean, and that 
the empty slots contain filler panels. For more information, refer to the 

Maintain Forced-Air 

Cooling Note to Users

 document available at 

ni.com/manuals

.

NI PXIe-4353 Calibration Procedure

This section provides instructions for verifying the voltage accuracy performance of the NI PXIe-4353. 
The calibration process consists of the following steps:

1.

Initial Setup for the NI PXIe-4353

—Install the module and configure it in Measurement & 

Automation Explorer (MAX).

2.

Verifying Module Operating Temperature Changes

—Verify that the current module operating 

temperature will not cause you to incorrectly calibrate your module.

3.

Verification Procedure

—Verify the existing operation of the module. This steps confirms whether 

the module is operating within its specified range and whether it needs adjustment.

4.

Adjustment

—If the module does not fall within the desired specifications, perform the adjustment 

procedure. This will adjust the module calibration constants. If the module does fall within the 
desired specifications, but the module has reached its calibration time interval since the last 
calibration adjustment, perform the adjustment procedure. Otherwise, performing the adjustment 
procedure is optional.

5.

Reverification

—Repeat the verification procedure to ensure that the module is operating within the 

published specifications after adjustment.

The first four steps are explained in the following sections.

1   

Refer to the Fluke 5520A user manual for the latest set up requirements and recommendations to achieve the best accuracy.

Summary of Contents for PXIe-4353

Page 1: ...ts 13 Module Verification Test Limit Tables 13 Where to Go for Support 15 Software Install NI DAQmx 9 3 or later on the calibration computer NI DAQmx includes high level function calls to simplify the...

Page 2: ...u use the instruments in Table 1 for calibrating an NI PXIe 4353 module Table 1 Recommended Equipment Equipment Recommended Model Requirements Voltage calibrator Fluke 5520A If this instrument is unav...

Page 3: ...re This section provides instructions for verifying the voltage accuracy performance of the NI PXIe 4353 The calibration process consists of the following steps 1 Initial Setup for the NI PXIe 4353 In...

Page 4: ...ration adjustment can cause you to incorrectly calibrate your module The onboard temperature sensor on the module can be used to verify that this condition does not exist Complete the following steps...

Page 5: ...4353 while referring to Figure 1 Figure 1 CAL 4353 to Fluke 5520A Calibrator Connections 2 Open a DAQmx calibration session on your device 3 Call the DAQmx Connect SC Express Calibration Accessory Cha...

Page 6: ...els you want to verify 7 Set the autozero mode to Every Sample and disable open thermocouple detection 8 On the DAQmx Channel property node set the DAQmx ADC timing mode property to Custom and set the...

Page 7: ...e of the analog input offset of the module 1 Open a DAQmx calibration session on your device 2 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the...

Page 8: ...bration session on your device 4 Call the DAQmx Connect SC Express Calibration Accessory Channels VI Configure the CAL 4353 by setting the connection to _cjc_calibration and the physical channels to t...

Page 9: ...Limit and Lower Limit values in Table 7 If the result is between these values the channel passes the test 13 Clear the acquisition 14 Repeat steps 5 through 12 for all CJC channels and test point valu...

Page 10: ...mx Connect SC Express Calibration Accessory Channels VI Set the connection to _tc_calibration and the physical channels to Dev1 ai0 31 4 Set the calibrator to output 70 mV and wait for the calibrator...

Page 11: ...0 Set the resistance calibrator to 2 wire compensation mode 11 Configure the CAL 4353 using the DAQmx Connect SC Express Calibration Accessory Channels VI Set the connection to _cjc_calibration and th...

Page 12: ...ose External Calibration VI and set the action to Commit This VI also saves the date time and temperature of the adjustment to the onboard memory Note If an error occurs during adjustment no constants...

Page 13: ...its Contains the Upper Limits and Lower Limits for the test point value That is when the module is within its 1 year calibration interval the test point value should fall between these upper and lower...

Page 14: ...02 3 0 000002 0 000002 4 0 000003 0 000003 5 0 000005 0 000005 6 0 0000065 0 0000065 7 High Speed 0 0000075 0 0000075 Table 6 NI PXIe 4353 Offset with Autozero Disabled Verification Test Limits Timing...

Page 15: ...side the United States visit the Worldwide Offices section of ni com niglobal to access the branch office Web sites which provide up to date contact information support phone numbers email addresses a...

Page 16: ...countries Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the approp...

Reviews: