background image

Table 1. Front Panel Connectors

Description

 

 

A

Access Status LED

B

Voltage Status LED

C

Chassis ground

D

Output LO

E

Sense LO

F

Triaxial connector pin Output HI

G

Triaxial connector inner shield, Guard

H

Triaxial connector pin, Sense HI

I

Triaxial connector outer shield, Chassis ground

J

Safety Interlock Input

K

Safety Interlock Pass Thru - Input

L

Safety Interlock Ground

M

Safety Interlock Pass Thru - Ground

Accessories

Table 2. PXIe-4135 Accessories

Accessory

Manufacturer

Part Number

 

 

 

Additional Connector Kit with Interlock Connector for
PXIe-4135 SMU

NI

784484-01

Safety Interlock Connector

Phoenix Contact 1790292

Configuring the PXIe-4135 in MAX

Use Measurement & Automation Explorer (MAX) to configure your NI hardware. MAX
informs other programs about which NI hardware products are in the system and how they are
configured. MAX is automatically installed with NI-DCPower.
1.

Launch MAX.

2.

In the configuration tree, expand 

Devices and Interfaces

 to see the list of installed NI

hardware.

12

  | 

ni.com

  | 

PXIe-4135 Getting Started Guide

Summary of Contents for PXIe-4135

Page 1: ...the product is damaged return it to NI for repair Contents Electromagnetic Compatibility Guidelines 2 Safety Guidelines for Hazardous Voltages 2 Verifying the System Requirements 2 Unpacking the Kit 3...

Page 2: ...sories connected to the LO Sense LO Output HI and Sense HI connectors Only twisted shielded pair cable i e Belden 83321E should be used to connect the signals of the LO Sense LO connector Any compatib...

Page 3: ...unding strap or by holding a grounded object such as your computer chassis 1 Touch the antistatic package to a metal part of the computer chassis 2 Remove the device from the package and inspect the d...

Page 4: ...tems not included in your device kit that you need to operate the PXIe 4135 Your application may require additional items not included in your kit to install or operate your device Required Items A PX...

Page 5: ...ree 2 Indoor use only Note Refer to the device specifications on ni com manuals for complete specifications Installing the Software You must be an Administrator to install NI software on your computer...

Page 6: ...that it is properly connected to safety ground Ensure that the PXIe 4135 is properly secured to the chassis using the two front panel mounting screws Double insulate all electrical connections that a...

Page 7: ...ed in series so that a single switch failure does not compromise safety protections Isolate switches so the operator cannot trigger or bypass the switches without the use of a tool Ensure the switches...

Page 8: ...talling the PXIe 4135 Caution To prevent damage to the PXIe 4135 caused by ESD or contamination handle the module using the edges or the metal bracket 1 Ensure the AC power source is connected to the...

Page 9: ...e 11 Cover all empty slots using EMC filler panels or fill using slot blockers to maximize cooling air flow depending on your application 12 Prepare the LO Sense LO connector assembly Caution To ensur...

Page 10: ...ngth for the safety interlock cable is 7 5 mm 0 295 in minimum and 10 mm 0 394 in absolute maximum The acceptable AWG range for the safety interlock cable is 16 24 b Use an insulation strip tool to ex...

Page 11: ...l ACCESS VOLTAGE LO HI GUARD GUARD INT PASS THRU SENSE LO SENSE HI SAFETY INTERLOCK 3 3 V MAX OUTPUT 200 V 1 A 3 A Pulse MAX 250 V MAX to PXIe 4135 Precision System SMU A C D E B K M J L F H I G PXIe...

Page 12: ...ries Table 2 PXIe 4135 Accessories Accessory Manufacturer Part Number Additional Connector Kit with Interlock Connector for PXIe 4135 SMU NI 784484 01 Safety Interlock Connector Phoenix Contact 179029...

Page 13: ...sure that the safety interlock input on the test fixture is closed 3 Set the niDCPower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Voltage for the PXIe 4135 4 Set the voltage...

Page 14: ...tatus Indicator is red and that a hazardous voltage error message appears 11 On the error message dialog click OK to prompt the PXIe 4135 to attempt to clear the error and re initialize the session to...

Page 15: ...quisition and measurement options using LabVIEW VIs or LabWindows CVI functions C or LabWindows CVI Available at Program Files IVI Foundation IVI Drivers NI DCPower Microsoft Visual C C NI DCPower doe...

Page 16: ...st with the PXI Express power rails a hardware module or the LED 1 Caution Apply external signals only while the PXIe 4135 is powered on Applying external signals while the module is powered off may c...

Page 17: ...te is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance fr...

Page 18: ...oducts technology refer to the appropriate location Help Patents in your software the patents txt file on your media or the National Instruments Patent Notice at ni com patents You can find informatio...