g)

Use the cable tie to provide strain relief for the cable to the backshell.

h)

Close the LO/sense LO connector assembly and tighten the retention screw to hold it
in place.

13. Attach output connections.

a)

Connect the triaxial cables to the module.

b)

Connect the LO/sense LO connector assembly to the module. Tighten the
thumbscrew on the connector assembly to hold it in place.

14. Ensure that the safety interlock connector is wired to a test fixture that ensures operator

safety, and prepare the safety interlock cable for insertion into the safety interlock
connector.
a)

Measure and mark your strip length on the safety interlock cable.

Note

 

The required wire strip length for the safety interlock cable is

7.5 mm (0.295 in.) minimum and 10 mm (0.394 in.) absolute maximum.
The acceptable AWG range for the safety interlock cable is 16-24.

b)

Use an insulation strip tool to expose cable of the appropriate length.

c)

The safety interlock connector accepts both solid and multi-strand conductor
cabling. If you are using a multi-stranded cable, twist the strands together before
insertion. For additional cabling reliability, strip and tin multi-stranded conductors
before insertion.

d)

Insert the cable.

e)

Inspect for loose strands and tighten any retention screws on the safety interlock
connector assembly to hold it in place.

f)

Connect the safety interlock connector to the device.

15. Power on the chassis.
16. Perform the safety interlock test.

Related Information

Testing the Safety Interlock

 on page 13

10

  | 

ni.com

  | 

PXIe-4135 Getting Started Guide

Summary of Contents for PXIe-4135

Page 1: ...the product is damaged return it to NI for repair Contents Electromagnetic Compatibility Guidelines 2 Safety Guidelines for Hazardous Voltages 2 Verifying the System Requirements 2 Unpacking the Kit 3...

Page 2: ...sories connected to the LO Sense LO Output HI and Sense HI connectors Only twisted shielded pair cable i e Belden 83321E should be used to connect the signals of the LO Sense LO connector Any compatib...

Page 3: ...unding strap or by holding a grounded object such as your computer chassis 1 Touch the antistatic package to a metal part of the computer chassis 2 Remove the device from the package and inspect the d...

Page 4: ...tems not included in your device kit that you need to operate the PXIe 4135 Your application may require additional items not included in your kit to install or operate your device Required Items A PX...

Page 5: ...ree 2 Indoor use only Note Refer to the device specifications on ni com manuals for complete specifications Installing the Software You must be an Administrator to install NI software on your computer...

Page 6: ...that it is properly connected to safety ground Ensure that the PXIe 4135 is properly secured to the chassis using the two front panel mounting screws Double insulate all electrical connections that a...

Page 7: ...ed in series so that a single switch failure does not compromise safety protections Isolate switches so the operator cannot trigger or bypass the switches without the use of a tool Ensure the switches...

Page 8: ...talling the PXIe 4135 Caution To prevent damage to the PXIe 4135 caused by ESD or contamination handle the module using the edges or the metal bracket 1 Ensure the AC power source is connected to the...

Page 9: ...e 11 Cover all empty slots using EMC filler panels or fill using slot blockers to maximize cooling air flow depending on your application 12 Prepare the LO Sense LO connector assembly Caution To ensur...

Page 10: ...ngth for the safety interlock cable is 7 5 mm 0 295 in minimum and 10 mm 0 394 in absolute maximum The acceptable AWG range for the safety interlock cable is 16 24 b Use an insulation strip tool to ex...

Page 11: ...l ACCESS VOLTAGE LO HI GUARD GUARD INT PASS THRU SENSE LO SENSE HI SAFETY INTERLOCK 3 3 V MAX OUTPUT 200 V 1 A 3 A Pulse MAX 250 V MAX to PXIe 4135 Precision System SMU A C D E B K M J L F H I G PXIe...

Page 12: ...ries Table 2 PXIe 4135 Accessories Accessory Manufacturer Part Number Additional Connector Kit with Interlock Connector for PXIe 4135 SMU NI 784484 01 Safety Interlock Connector Phoenix Contact 179029...

Page 13: ...sure that the safety interlock input on the test fixture is closed 3 Set the niDCPower Output Function property or NIDCPOWER_OUTPUT_FUNCTION attribute to DC Voltage for the PXIe 4135 4 Set the voltage...

Page 14: ...tatus Indicator is red and that a hazardous voltage error message appears 11 On the error message dialog click OK to prompt the PXIe 4135 to attempt to clear the error and re initialize the session to...

Page 15: ...quisition and measurement options using LabVIEW VIs or LabWindows CVI functions C or LabWindows CVI Available at Program Files IVI Foundation IVI Drivers NI DCPower Microsoft Visual C C NI DCPower doe...

Page 16: ...st with the PXI Express power rails a hardware module or the LED 1 Caution Apply external signals only while the PXIe 4135 is powered on Applying external signals while the module is powered off may c...

Page 17: ...te is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance fr...

Page 18: ...oducts technology refer to the appropriate location Help Patents in your software the patents txt file on your media or the National Instruments Patent Notice at ni com patents You can find informatio...

Reviews: