Figure 8. PXIe-1092 Vents
PXI
e-1092
3
4
5
7
6
2
DISCONNECT PO
WER
CORD BEFORE SER
VICING.
1
1. PXI Module Air Intake (2x)
2. Power Supply Intake
3. PXI Module Air Exhaust Vent
4. Power Supply Air Exhaust Vent
5. Timing and Synchronization Upgrade Air Exhaust
Vent
6. Timing and Synchronization Upgrade Air Intake
7. Side Air Intake Vent (Right)/Side Air Exhaust Vent
(Left)
Note
The side exhaust vent (not shown) is located on the left side of the chassis.
Chassis Ambient Temperature Definition
The chassis fan control system uses ambient intake air temperatures for controlling fan speeds
when in Auto mode. These temperatures may be higher than ambient room temperature
depending on surrounding equipment and/or blockages. Ensure ambient intake temperatures
do not exceed the ratings in the
Operating Environment
section of the
PXIe-1092
PXIe-1092 User Guide
|