NI 5122/5124/5142 Calibration Procedure
|
© National Instruments
|
9
4.
Configure the horizontal properties using the niScope Configure Horizontal Timing VI.
5.
Commit all the parameter settings to hardware using the niScope Commit VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureVerti
cal
with the following
parameters:
coupling
:
NISCOPE_VAL_DC
probeAttenuation
:
1.0
vi
: The instrument handle from
niScope_init
channelList
:
"0"
range
: The
Range
value in
Table 3 for the current iteration
offset
:
0.0
enabled
:
NISCOPE_VAL_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_Configure
HorizontalTiming
with the
following parameters:
enforceRealtime
:
NISCOPE_VAL_TRUE
numRecords
:
1
vi
: The instrument handle from
niScope_init
minSampleRate
:
10,000,000
refPosition
:
50.0
minNumPts
:
100,000
LabVIEW VI
C/C++ Function Call
Call
niScope_Commit
with the
following parameter:
vi
: The instrument handle from
niScope_init