NI 5122/5124/5142 Calibration Procedure
|
© National Instruments
|
31
You have finished verifying the bandwidth and flatness of the NI 5122/5124/5142.
LabVIEW VI
C/C++ Function Call
Call
niScope_close
with the
following parameter:
vi
: The instrument handle from
niScope_init
Table 12.
NI 5122/5142 Bandwidth and Flatness Stimuli
and Published Specifications
Iteration
Max Input
Frequency
Input
Impedance
Range
(V)
Input
Frequency
(Hz)
NI 5122/5142
Published
Specifications
Max
Level
(dB)
Min
Level
(dB)
1
100 MHz
50
Ω
0.2
20,100,000
0.4
-0.4
40,100,000
1
-1
80,100,000
3
†
-3
†
2
100 MHz
50
Ω
0.4
20,100,000
0.4
-0.4
50,100,000
1
-1
100,100,000
3
†
-3
†
3
100 MHz
1 M
Ω
0.2
20,100,000
0.4
-0.4
40,100,000
1
-1
80,100,000
3
†
-3
†
4
100 MHz
1 M
Ω
0.4
20,100,000
0.4
-0.4
50,100,000
1
-1
100,100,000
3
†
-3
†