NI 5122/5124/5142 Calibration Procedure
Vertical Offset Accuracy
Complete the following steps to verify vertical offset accuracy of the NI 5122/5124/5142. You
must verify both channels with each iteration in Table 3.
1.
Open a session and obtain a session handle using the niScope Initialize VI.
Note
Throughout the procedure, refer to the C/C++ function call parameters for the
LabVIEW input values.
2.
Configure the input impedance and the maximum input frequency using the
niScope Configure Chan Characteristics VI.
3.
Configure the common vertical properties using the niScope Configure Vertical VI.
LabVIEW VI
C/C++ Function Call
Call
niScope_init
with the
following parameters:
resourceName
: The device name
assigned by MAX
idQuery
:
VI_FALSE
resetDevice
:
VI_TRUE
LabVIEW VI
C/C++ Function Call
Call
niScope_ConfigureChan
Characteristics
with the
following parameters:
vi
: The instrument handle from
niScope_init
channelList
:
"0"
inputImpedance
: The
Input
Impedance
value in Table 3 for
the current iteration
maxInputFrequency
: The
Maximum Input Frequency
value in Table 3 for the current
iteration