NI 5122/5124/5142 Calibration Procedure
|
© National Instruments
|
11
12. End the session using the niScope Close VI.
You have finished verifying the vertical offset accuracy of the NI 5122/5124/5142.
LabVIEW VI
C/C++ Function Call
Call
niScope_close
with the
following parameter:
vi
: The instrument handle from
niScope_init
Table 4.
NI 5122 Vertical Offset Limits
Range
NI PXI/PXIe-5122
NI PCI-5122
Calibration
Test Limits
(V)
Published
Specifications
(V)
Calibration
Test Limits
(V)
Published
Specifications
(V)
0.2
±0.00058
±0.001
±0.00114
±0.002
0.4
±0.00056
±0.001
±0.00114
±0.002
1
±0.00076
±0.0012
±0.00114
±0.002
2
±0.0008
±0.0016
±0.00114
±0.002
4
±0.0033
±0.008
±0.0033
±0.008
10
±0.0036
±0.008
±0.0036
±0.008
20
±0.005
±0.013
±0.005
±0.013
Table 5.
NI 5124 Vertical Offset Limits
Range
NI PXI-5124
NI PCI-5124
Calibration
Test Limits
(V)
Published
Specifications
(V)
Calibration
Test Limits
(V)
Published
Specifications
(V)
0.2
±0.0008
±0.0013
±0.0013
±0.0018
0.4
±0.0008
±0.0013
±0.0013
±0.0018
1
±0.0011
±0.0015
±0.0016
±0.0021
2
±0.0011
±0.0015
±0.0016
±0.0021
4
±0.00575
±0.01
±0.005
±0.010