NI 5122/5124/5142 Calibration Procedure
|
© National Instruments
|
7
18
50
Ω
20,000,000
20,000,000
10
19
1 M
Ω
100,000,000
150,000,000
0.2
20
1 M
Ω
100,000,000
150,000,000
0.4
21
1 M
Ω
100,000,000
150,000,000
1
22
1 M
Ω
100,000,000
150,000,000
2
23
1 M
Ω
100,000,000
150,000,000
4
24
1 M
Ω
100,000,000
150,000,000
10
25
1 M
Ω
100,000,000
150,000,000
20
26
1 M
Ω
35,000,000
60,000,000
0.2
27
1 M
Ω
35,000,000
60,000,000
0.4
28
1 M
Ω
35,000,000
60,000,000
1
29
1 M
Ω
35,000,000
60,000,000
2
30
1 M
Ω
35,000,000
60,000,000
4
31
1 M
Ω
35,000,000
60,000,000
10
32
1 M
Ω
35,000,000
60,000,000
20
33
1 M
Ω
20,000,000
20,000,000
0.2
34
1 M
Ω
20,000,000
20,000,000
0.4
35
1 M
Ω
20,000,000
20,000,000
1
36
1 M
Ω
20,000,000
20,000,000
2
37
1 M
Ω
20,000,000
20,000,000
4
38
1 M
Ω
20,000,000
20,000,000
10
39
1 M
Ω
20,000,000
20,000,000
20
Table 3.
NI 5122/5124/5142 Input Parameters for Vertical Offset Accuracy
and Vertical Gain Accuracy Verification (Continued)
Iteration
Input
Impedance
Max Input Frequency (Hz)
Range
NI 5122/5142
NI 5124