NI 5122/5124/5142 Calibration Procedure
Vertical Offset and Vertical Gain Accuracy
Table 3 contains the input parameters for verifying both vertical offset accuracy and vertical
gain accuracy of the digitizer.
To verify vertical offset accuracy, complete the procedures described in the
section for each of the 39 iterations in Table 3 for channel 0, then repeat the procedures
for channel 1. The
Calibration Test Limits
and
Published Specifications
for vertical offset
accuracy are shown in Table 4 (NI 5122), Table 5 (NI 5124), and Table 6 (NI 5142).
To verify vertical gain accuracy, complete the procedures described in the
section for each of the 39 iterations in Table 3 for channel 0, then repeat the procedures
for channel 1. The
Calibration Test Limits
and
Published Specifications
for vertical gain
accuracy are shown in Table 7 (NI 5122/5142) and Table 8 (NI 5124).
Table 3.
NI 5122/5124/5142 Input Parameters for Vertical Offset Accuracy
and Vertical Gain Accuracy Verification
Iteration
Input
Impedance
Max Input Frequency (Hz)
Range
NI 5122/5142
NI 5124
1
50
Ω
100,000,000
150,000,000
0.2
2
50
Ω
100,000,000
150,000,000
0.4
3
50
Ω
100,000,000
150,000,000
1
4
50
Ω
100,000,000
150,000,000
2
5
50
Ω
100,000,000
150,000,000
4
6
50
Ω
100,000,000
150,000,000
10
7
50
Ω
35,000,000
60,000,000
0.2
8
50
Ω
35,000,000
60,000,000
0.4
9
50
Ω
35,000,000
60,000,000
1
10
50
Ω
35,000,000
60,000,000
2
11
50
Ω
35,000,000
60,000,000
4
12
50
Ω
35,000,000
60,000,000
10
13
50
Ω
20,000,000
20,000,000
0.2
14
50
Ω
20,000,000
20,000,000
0.4
15
50
Ω
20,000,000
20,000,000
1
16
50
Ω
20,000,000
20,000,000
2
17
50
Ω
20,000,000
20,000,000
4