Table 7 describes the signal connections for the NI 5160/5162 (4 CH).
NI 5185/5186 Front Panels
Figure 20 shows the NI 5185/5186 front panels.
Figure 20.
NI 5185/5186 Front Panels
Table 7.
NI 5160/5162 (4 CH)
Front Panel Signal Connections
Connector
Description
Function
CH 0 ... CH 3
Standard BNC connector
Analog input connection; digitizes data and
triggers acquisitions.
CLK IN
Standard SMB jack
Reference clock in, sample clock in.
CLK OUT
Standard SMB jack
Reference clock out.
PFI 0
Standard SMB jack
Digital trigger in/out.
PFI 1
Standard SMB jack
Digital trigger in/out, probe compensation.
5 GHz 12.5 GS/s 8-Bit Digitizer
NI PXIe-5186
+16 dBm MAX
(1.4 Vrms)
50
Ω
±5 V
MAX
50
Ω
50
Ω
1M
Ω
+9 dBm MAX
(0.6 Vrms)
50
Ω
50
Ω
1M
Ω
+16 dBm MAX
(1.4 Vrms)
50
Ω
±5 V
MAX
50
Ω
50
Ω
1M
Ω
+9 dBm MAX
(0.6 Vrms)
50
Ω
50
Ω
1M
Ω
3 GHz 12.5 GS/s 8-Bit Digitizer
NI PXIe-5185
1 Vpp measure
1 Vpp measure
10 Vpp measure
10 Vpp measure
± 1 V MAX
± 1 V MAX
± 42 V MAX
± 42 V MAX
1 Vpp measure
1 Vpp measure
10 Vpp measure
10 Vpp measure
± 1 V MAX
± 1 V MAX
± 42 V MAX
± 42 V MAX