CALIBRATION PROCEDURE

FD-11613/11614

8-Channel/16-Channel Temperature Input Device for FieldDAQ

This document contains information for calibrating the FD-11613 and FD-11614. For more
information about calibration, visit 

ni.com/calibration

.

Contents

Software Requirements............................................................................................................. 1
Documentation.......................................................................................................................... 2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................2
Calibration Procedure............................................................................................................... 3

Initial Setup....................................................................................................................... 3
Verification........................................................................................................................4
Adjustment........................................................................................................................ 7
EEPROM Update..............................................................................................................8
Reverification....................................................................................................................9

Accuracy Under Calibration Conditions...................................................................................9
Where to Go for Support.........................................................................................................10

Software Requirements

Calibrating the FD-11613 or FD-11614 requires the installation of NI-DAQmx on the
calibration system. NI recommends using the latest NI-DAQmx driver. Earliest driver support
version for calibrating the FD-11613 or FD-11614 is listed in the following table.

Table 1. FD-11613/11614 Driver Support

Driver

Earliest Version Support for Device Calibration

 

 

NI-DAQmx

18.1

You can download NI-DAQmx from 

ni.com/downloads

. NI-DAQmx supports many

programming languages, including LabVIEW, LabWindows

/CVI

, C/C++, C#, and Visual

Basic .NET. When you install NI-DAQmx, you only need to install support for the application
software that you intend to use.

Summary of Contents for FD-11613

Page 1: ...to Go for Support 10 Software Requirements Calibrating the FD 11613 or FD 11614 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI DAQmx driver Earliest...

Page 2: ...pplications that use the NI DAQmx driver NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties Test Equipment The following table lists the equipment requi...

Page 3: ...rt Note You must install NI DAQmx 18 1 or later for device calibration support 2 Set up the FieldDAQ device as outlined in the FD 11613 11614 Quick Start 3 Configure the FieldDAQ device in Measurement...

Page 4: ...of operation and test points required to verify the FieldDAQ device The verification procedures assume that adequate traceable uncertainties are available for the calibration references Complete the...

Page 5: ...Accuracy Under Calibration Conditions 4 Set the calibrator to Operate mode OPR 5 Acquire and average samples a Create and configure an AI voltage channel on the FieldDAQ device according to the follow...

Page 6: ...om each channel and average the readings e Clear the task 13 Compare the average to the limits in the following table Table 6 FD 11613 11614 Voltage Verification Test Limits for Zero Test Points ADC T...

Page 7: ...eldDAQ device as shown in the following figure Figure 3 Bank 1 Voltage Channel Adjustment Connections TC0 TC0 HI LO TC1 TC1 TC6 TC6 TC7 TC7 Calibrator FD 11613 11614 TC2 TC2 TC3 TC3 TC4 TC4 TC5 TC5 Ba...

Page 8: ...gh g for each reference value in the array of adjustment points 4 Disconnect the calibrator from the FieldDAQ device 5 FD 11614 Connect the calibrator to Bank 2 of the FD 11614 as shown in the followi...

Page 9: ...onditions The values in the following table are based on calibrated scaling coefficients which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the followin...

Page 10: ...d Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents coverin...

Reviews: