The following accuracy table is valid for calibration under the following conditions:

Ambient temperature 23 °C ± 5 °C

No nodes share power with the FieldDAQ device under calibration

The test limits listed in Table 3 are derived using the values in the following table.

Table 7. FD-11603 Voltage Accuracy Under Calibration Conditions

Percent of Reading (Gain Error)

Percent of Range (Offset)

 

 

±0.037%

±1.05 mV

Note

 

Values in this table are intended for calibration verification only. These

values only apply under the specific calibration conditions described in this
document, and are not to be construed as general operational specifications of the
FD-11603. For operational specifications, refer to the most recent 

FD-11603

Specifications

 at 

ni.com/manuals

.

Signal Connector Pinout

Figure 6. FD-11603 Pinout

AI–

NC

NC

AI+

NC

2

3

4

5

1

Table 8. Signal Descriptions

Pin Number

Wire Color*

Signal

Description

 

 

 

 

1

3

5

Brown

Blue

Gray

NC

No connect. Do not connect signals to
this terminal.

2

White

AI-

Negative analog input voltage signal

4

Black

AI+

Positive analog input voltage signal

* Wire color pertains to SHM125M I/O cables sold through NI. Other manufacturers' cable
wire colors may vary.

FD-11603 Calibration Procedure

  | 

© National Instruments

  | 

9

Summary of Contents for FD-11603

Page 1: ...inout 9 Where to Go for Support 10 Software Requirements Calibrating the FD 11603 requires the installation of NI DAQmx on the calibration system NI recommends using the latest NI DAQmx driver Earlies...

Page 2: ...ovides detailed specifications for your FieldDAQ device NI DAQmx Readme Provides operating system and application software support in NI DAQmx NI DAQmx Help Provides information about creating applica...

Page 3: ...d wires act as antennas picking up extra noise that can affect measurements Verify that all connections to the device are secure Use shielded copper wire for all cable connections to the device Use tw...

Page 4: ...ng the Device in MAX When the FieldDAQ device is connected to a network multiple users can access the device To perform any DAQ functionality on the device including reset and self test you must reser...

Page 5: ...alue determined by the array of recommended calibration voltages indicated in the following table Table 3 FD 11603 Test Limits Test Point 1 Year Limits Location Value V Lower Limit V Upper Limit V Neg...

Page 6: ...samples e Average the readings f Clear the task 6 Set the calibrator to Standby mode STBY 7 Compare the average to the limits in Table 3 8 Repeat steps 3 through 7 for each test point 9 Disconnect the...

Page 7: ...n degrees Celsius using the Set Temperature FieldDAQ function b Call the Get FD 11603 Calibration Adjustment Points function c Set the calibrator to a reference value determined by the array of adjust...

Page 8: ...Initialize External Calibration function to start a calibration session on the FieldDAQ device The default password is NI 2 Call the DAQmx Set Temperature FieldDAQ Calibration function to input the ex...

Page 9: ...fic calibration conditions described in this document and are not to be construed as general operational specifications of the FD 11603 For operational specifications refer to the most recent FD 11603...

Page 10: ...nd Logo Guidelines at ni com trademarks for information on NI trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents coveri...

Reviews: