background image

Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at 

ni.com/trademarks

 for

information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their
respective companies. For patents covering NI products/technology, refer to the appropriate location: 

Help

»

Patents in your

software, the 

patents.txt

 file on your media, or the National Instruments Patent Notice at 

ni.com/patents

. You can find

information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer
to the Export Compliance Information at 

ni.com/legal/export-compliance

 for the NI global trade compliance policy and how

to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS
TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S.
Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable
limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.

© 2017—2020 National Instruments Corporation. All rights reserved.

378039C-01

September 2, 2020

Summary of Contents for C Series

Page 1: ...fication 6 Accuracy under Calibration Conditions 6 Revision History 7 Worldwide Support and Services 7 Software Calibrating the NI 9202 requires the installation of NI DAQmx 17 6 or later on the calib...

Page 2: ...DAQmx VIs and functions NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties NI DAQmx NET Help Support for Visual Studio Reference information for NI DAQm...

Page 3: ...airs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the device measu...

Page 4: ...ure 1 Analog Input Verification Connections NI 9202 Calibrator AIx AIx COM HI LO Guard NI 9202 Calibrator AIx AIx COM HI LO 1 2 1 Connections when using a calibrator with a guard connection 2 Connecti...

Page 5: ...est Limits Test Points 2 Year Limits Location Value VDC Lower Limit VDC Upper Limit VDC Min 10 0000 10 0298 9 9701 Mid 0 0000 0 0158 0 0158 Max 10 0000 9 9701 10 0298 Note The test limits listed in th...

Page 6: ...he calibration session j Clear the task 4 Disconnect the NI 9202 from the calibrator EEPROM Update When an adjustment procedure is completed the NI 9202 internal calibration memory EEPROM is immediate...

Page 7: ...ete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Applicatio...

Page 8: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to...