The following accuracy table is valid for calibration under the following conditions:
•
Ambient temperature 23 °C ± 5 °C
•
NI 9202 installed in slot 8 of a cDAQ-9178 chassis
•
Slots 1 through 7 of the cDAQ-9178 chassis are empty
Table 6. NI 9202 Accuracy under Calibration Conditions
Device
Percent of Reading (Gain Error)
Percentage of Range (Offset Error)
2
NI 9202
0.14%
0.15%
Note
For operational specifications, refer to the most recent
NI 9202 Datasheet
.
Revision History
Part Number Edition Date
Section
Changes
378039C-01 August 2020 Accuracy Verification
Updated the AI voltage task table
Updated Channel Property setting
sequence in step 5.c)
Updated the verification test limits
table
Worldwide Support and Services
The NI website is your complete resource for technical support. At
access to everything from troubleshooting and application development self-help resources to
email and phone assistance from NI Application Engineers.
Visit
for information about the services NI offers.
Visit
to register your NI product. Product registration facilitates technical
support and ensures that you receive important information updates from NI.
NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas,
78759-3504. NI also has offices located around the world. For support in the United States,
create your service request at
or dial 1 866 ASK MYNI (275 6964). For
support outside the United States, visit the
Worldwide Offices
section of
to
access the branch office websites, which provide up-to-date contact information.
2
Range equals 10.58 V.
NI 9202 Calibration Procedure
|
© National Instruments Corporation
|
7