Table 3. NI 9202 Task Timing

Samples Per Channel

Acquisition Mode

Rate (kS/s)

 

 

 

1000

Finite Samples

10

c)

Set the filter frequency to 625 Hz using the Channel Property ( 

Analog Input

 »

General Properties

 »

Filter

 »

Frequency

 ).

d)

Start the task.

e)

Read the samples from the NI 9202.

f)

Average the samples.

g)

Clear the task.

6.

Compare the averaged samples to the limits in the 

NI 9202 Verification Test Limits

 table

to verify the input accuracy. If the values are within these limits, the device passes the
test.

7.

Set the calibrator to Standby mode (STBY).

8.

Repeat steps 3 through 7 for each test point in the 

NI 9202 Verification Test Limits

 table.

9.

Disconnect the calibrator from the device.

Table 4. NI 9202 Verification Test Limits

Test Points

2-Year Limits

Location

Value (VDC)

Lower Limit (VDC)

Upper Limit (VDC)

 

 

 

 

Min

-10.0000

-10.0298

-9.9701

Mid

0.0000

-0.0158

0.0158

Max

10.0000

9.9701

10.0298

Note

 

The test limits listed in the 

NI 9202 Verification Test Limits

 table are derived

using values from the 

NI 9202 Accuracy Under Calibration Conditions

 table.

Adjustment

The following performance adjustment procedure describes the sequence of operation required
to adjust the NI 9202.

Accuracy Adjustment

Complete the following procedure to adjust the accuracy performance of the NI 9202.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9202 to the calibrator as shown in the 

Accuracy Verification

 section.

3.

Adjust the NI 9202.
a)

Initialize a calibration session on the NI 9202. The default password is 

NI

.

b)

Input the external temperature in degrees Celsius.

NI 9202 Calibration Procedure

  | 

© National Instruments Corporation

  | 

5

Summary of Contents for C Series

Page 1: ...fication 6 Accuracy under Calibration Conditions 6 Revision History 7 Worldwide Support and Services 7 Software Calibrating the NI 9202 requires the installation of NI DAQmx 17 6 or later on the calib...

Page 2: ...DAQmx VIs and functions NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties NI DAQmx NET Help Support for Visual Studio Reference information for NI DAQm...

Page 3: ...airs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the device measu...

Page 4: ...ure 1 Analog Input Verification Connections NI 9202 Calibrator AIx AIx COM HI LO Guard NI 9202 Calibrator AIx AIx COM HI LO 1 2 1 Connections when using a calibrator with a guard connection 2 Connecti...

Page 5: ...est Limits Test Points 2 Year Limits Location Value VDC Lower Limit VDC Upper Limit VDC Min 10 0000 10 0298 9 9701 Mid 0 0000 0 0158 0 0158 Max 10 0000 9 9701 10 0298 Note The test limits listed in th...

Page 6: ...he calibration session j Clear the task 4 Disconnect the NI 9202 from the calibrator EEPROM Update When an adjustment procedure is completed the NI 9202 internal calibration memory EEPROM is immediate...

Page 7: ...ete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Applicatio...

Page 8: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to...

Reviews: