Table 3. NI 9202 Task Timing
Samples Per Channel
Acquisition Mode
Rate (kS/s)
1000
Finite Samples
10
c)
Set the filter frequency to 625 Hz using the Channel Property (
Analog Input
»
General Properties
»
Filter
»
Frequency
).
d)
Start the task.
e)
Read the samples from the NI 9202.
f)
Average the samples.
g)
Clear the task.
6.
Compare the averaged samples to the limits in the
NI 9202 Verification Test Limits
table
to verify the input accuracy. If the values are within these limits, the device passes the
test.
7.
Set the calibrator to Standby mode (STBY).
8.
Repeat steps 3 through 7 for each test point in the
NI 9202 Verification Test Limits
table.
9.
Disconnect the calibrator from the device.
Table 4. NI 9202 Verification Test Limits
Test Points
2-Year Limits
Location
Value (VDC)
Lower Limit (VDC)
Upper Limit (VDC)
Min
-10.0000
-10.0298
-9.9701
Mid
0.0000
-0.0158
0.0158
Max
10.0000
9.9701
10.0298
Note
The test limits listed in the
NI 9202 Verification Test Limits
table are derived
using values from the
NI 9202 Accuracy Under Calibration Conditions
table.
Adjustment
The following performance adjustment procedure describes the sequence of operation required
to adjust the NI 9202.
Accuracy Adjustment
Complete the following procedure to adjust the accuracy performance of the NI 9202.
1.
Set the calibrator to Standby mode (STBY).
2.
Connect the NI 9202 to the calibrator as shown in the
Accuracy Verification
section.
3.
Adjust the NI 9202.
a)
Initialize a calibration session on the NI 9202. The default password is
NI
.
b)
Input the external temperature in degrees Celsius.
NI 9202 Calibration Procedure
|