Table 1. Recommended Equipment

Equipment

Recommended Model

Requirements

 

 

 

Calibrator

Fluke 5500A

Use a high-precision voltage source
with a gain error of <57 ppm and an
offset error of <2 mV.

Chassis

cDAQ-9178 (part number
781156-01)

DSUB Connection
Accessory

1

NI 9923 (part number
781503-01)

Test Conditions

The following setup and environmental conditions are required to ensure the device meets
published specifications.

Keep connections to the device as short as possible. Long cables and wires act as
antennas, picking up extra noise that can affect measurements.

Verify that all connections to the device are secure.

Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire
to eliminate noise and thermal offsets.

Maintain an ambient temperature of 23 °C ± 5 °C.

Keep relative humidity below 80%.

Allow a warm-up time of at least 10 minutes to ensure that the device measurement
circuitry is at a stable operating temperature.

Initial Setup

Complete the following steps to set up the NI 9202.
1.

Install NI-DAQmx.

2.

Make sure that the cDAQ-9178 power source is not connected to the chassis.

3.

Connect the cDAQ-9178 chassis to the system safety ground.
a)

Attach a ring lug to a 14 AWG (1.6 mm) wire.

b)

Connect the ring lug to the ground terminal on the side of the cDAQ-9178 using the
ground screw.

c)

Attach the other end of the wire to the system safety ground.

4.

Install the module in slot 8 of the cDAQ-9178 chassis. Leave slots 1 through 7 of the
cDAQ-9178 chassis empty.

1

For NI 9202 with DSUB variant

NI 9202 Calibration Procedure

  | 

© National Instruments Corporation

  | 

3

Summary of Contents for C Series

Page 1: ...fication 6 Accuracy under Calibration Conditions 6 Revision History 7 Worldwide Support and Services 7 Software Calibrating the NI 9202 requires the installation of NI DAQmx 17 6 or later on the calib...

Page 2: ...DAQmx VIs and functions NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties NI DAQmx NET Help Support for Visual Studio Reference information for NI DAQm...

Page 3: ...airs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the device measu...

Page 4: ...ure 1 Analog Input Verification Connections NI 9202 Calibrator AIx AIx COM HI LO Guard NI 9202 Calibrator AIx AIx COM HI LO 1 2 1 Connections when using a calibrator with a guard connection 2 Connecti...

Page 5: ...est Limits Test Points 2 Year Limits Location Value VDC Lower Limit VDC Upper Limit VDC Min 10 0000 10 0298 9 9701 Mid 0 0000 0 0158 0 0158 Max 10 0000 9 9701 10 0298 Note The test limits listed in th...

Page 6: ...he calibration session j Clear the task 4 Disconnect the NI 9202 from the calibrator EEPROM Update When an adjustment procedure is completed the NI 9202 internal calibration memory EEPROM is immediate...

Page 7: ...ete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Applicatio...

Page 8: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to...

Reviews: