Connect the cDAQ-9178 chassis to your host computer.

6.

Connect the power source to the cDAQ-9178 chassis.

7.

Launch Measurement & Automation Explorer (MAX).

8.

Right-click the device name and select 

Self-Test

 to ensure that the module is working

properly.

Verification

The following performance verification procedure describes the sequence of operation and
provides test points required to verify the NI 9202. The verification procedure assumes that
adequate traceable uncertainties are available for the calibration references.

Accuracy Verification

Complete the following procedure to determine the As-Found status of the NI 9202.
1.

Set the calibrator to Standby mode (STBY).

2.

Connect the NI 9202 to the calibrator as shown in the following figure.

Figure 1. Analog Input Verification Connections

NI 9202

Calibrator

AIx

AIx+

COM

HI

LO

Guard

NI 9202

Calibrator

AIx

AIx+

COM

HI

LO

1

2

1. Connections when using a calibrator with a guard connection.
2. Connections when using a calibrator with no guard connection.

Note

 

If the calibrator outputs are truly floating, connect the negative output to

a quiet earth ground as well as COM to give the entire system a ground
reference.

3.

Set the calibrator voltage to a Test Point value indicated in the 

NI 9202 Verification Test

Limits

 table.

4.

Set the calibrator to Operate mode (OPR).

5.

Acquire and average samples.
a)

Create and configure an AI voltage channel according to the following 

NI 9202 AI

Voltage Task

 table.

Table 2. NI 9202 AI Voltage Task

Physical Channels

Range

Terminal Configuration

 

 

 

cDAQMod8/AI0:15

±10 V

Differential

b)

Configure the AI voltage channel timing according to the following 

NI 9202 Task

Timing

 table.

4

  | 

ni.com

  | 

NI 9202 Calibration Procedure

Summary of Contents for C Series

Page 1: ...fication 6 Accuracy under Calibration Conditions 6 Revision History 7 Worldwide Support and Services 7 Software Calibrating the NI 9202 requires the installation of NI DAQmx 17 6 or later on the calib...

Page 2: ...DAQmx VIs and functions NI DAQmx C Reference Help Reference information for NI DAQmx C functions and NI DAQmx C properties NI DAQmx NET Help Support for Visual Studio Reference information for NI DAQm...

Page 3: ...airs wire to eliminate noise and thermal offsets Maintain an ambient temperature of 23 C 5 C Keep relative humidity below 80 Allow a warm up time of at least 10 minutes to ensure that the device measu...

Page 4: ...ure 1 Analog Input Verification Connections NI 9202 Calibrator AIx AIx COM HI LO Guard NI 9202 Calibrator AIx AIx COM HI LO 1 2 1 Connections when using a calibrator with a guard connection 2 Connecti...

Page 5: ...est Limits Test Points 2 Year Limits Location Value VDC Lower Limit VDC Upper Limit VDC Min 10 0000 10 0298 9 9701 Mid 0 0000 0 0158 0 0158 Max 10 0000 9 9701 10 0298 Note The test limits listed in th...

Page 6: ...he calibration session j Clear the task 4 Disconnect the NI 9202 from the calibrator EEPROM Update When an adjustment procedure is completed the NI 9202 internal calibration memory EEPROM is immediate...

Page 7: ...ete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance from NI Applicatio...

Page 8: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy and how to...

Reviews: