Cleaning & maintenance
Clean the surface of JuLI™ FL instrument with a damp cloth.
To clean the LCD screen, turn off JuLI™ FL instrument, disconnect the power cable,
and clean the LCD screen with a soft cloth lightly moistened with LCD cleansing
detergent. Cleaning the screen with excessive force can damage the LCD screen.
If liquid spills on JuLI™ FL, turn off the power immediately and wipe dry.
JuLI™ FL does not need regular maintenance. To troubleshoot problems
with JuLI™ FL, contact Technical Support (Refer to page 72).
IMPORTANT
Never disassemble or service JuLI™ FL by yourself.
Unauthorized repairs may damage JuLI™ FL or alter its functionality,
which will void your warranty. Contact
or your
local JuLI™ FL distributor to arrange for service.
Always wipe surfaces with ethanol-soaked paper towels.
Do not directly spray ethanol anywhere on JuLI™ FL.
Avoid exposing JuLI™ FL to UV light
UV light may degrade components, including plastic. Damage from UV
exposure is not covered under the manufacturer’s warranty.