Set up parameter
Settings menu works for each scope unit if 2
nd
unit is connected to
the station unit. Press the
Channel tab
to select each scope unit.
Set the recommended parameter values for corresponding to cell
line. The default value is
A (Bright mode only).
Sensitivity :
Refers to the contrast of the objects from the background. The
increased sensitivity provides better recognition of fainter objects
in the Bright mode.
Background level:
The decreased background level provides better recognition of
fainter objects in Bright mode.
· Press the Protocol button to create the new protocol or edit existing one.
Cell line
HepG2, GH3, Hep3B,
A549, MCF7, SH-SY5Y,
SCN2.2, NIH-3T3, F9,
HeLa, DU-145
LNCaP
U-2 OS
ADSC
Recommendation
Sensitivity level 9
Background level 2
Sensitivity level 10
Background level 2
Sensitivity level 10
Background level 8
Parameter
A(Default)
B
C