Figure 7.
6. Unscrew the dew point sensor from the sensor block and
replace with the new or re-calibrated sensor. (See Figure 7)
7. Reattach the new or re-calibrated dew point sensor and sensor block
assembly to the controller chassis plate using the 2x fixing screws.
8. Replace the sensor plug and screw to complete the sensor assembly.
(See Figure 3 & 4)
9. Close the shroud and ensure the latches are in position.
10. Replace the top cover and insert the 4x M5 screws. These screws
should be hand tightened only. (See Figure 1)
DEW POINT SENSOR
SENSOR BLOCK
Figure 5.
5. Remove the 2x fixing screws from the controller chassis plate and
release the dew point sensor assembly. (See Figure 5 & 6)
When service C is complete reset the dryer, refer to:
Figure 6.
NOTE:
Allow sensor to stabilize, this may take several hours.
CH. 13. RESETTING DRYER CONTROLLER