  Electric power infrastructure controlled environments (i.e. step down transformers or high voltage power 

lines) 

Implanted Cardiac or Other Neurostimulation Systems

—Patients who have implanted cardiac or other 

neurostimulation systems should not use the Nalu Neurostimulation System. Electrical pulses from the Nalu 
Neurostimulation System may interact with the sensing operation of an implanted cardiac or neurostimulation 
system, causing the system to respond inappropriately. 

Warnings 

Electromagnetic Interference (EMI)

—EMI is a field of energy generated by equipment found in the home, work, 

medical, or public environments. Very strong EMI can interfere with the System. The device includes features that 
provide protection from EMI. Most electrical device and magnets encountered in a normal day will not affect the 
operation of the System, however, strong sources of EMI could result in the following:

 

  Serious user injury resulting from heating of the implanted device and damage to surrounding tissue. 

  System damage resulting in a loss of, or change in, symptom control that might require additional surgery. 

  Operational changes to the Therapy Disc. This may cause the external device to turn on, turn off, or reset 

to factory settings. If this occurs, the Therapy Disc needs to be reprogrammed. 

  Unexpected changes in stimulation, causing a momentary increase in stimulation or intermittent stimulation. 

Some users have described a jolting or shocking sensation. Although the unexpected change in stimulation 
may feel uncomfortable, it will not damage the device or cause direct injury to the user. In rare cases, as a 
result of the unexpected changes in stimulation, users have fallen down and been injured. 

If you suspect that your Nalu Neurostimulation System is being affected by EMI, then you should: 

 

For the Therapy Disc

 – Remove or Turn Off 

 

For the Trial Therapy Disc

 – Disconnect and/or Turn Off Stimulation 

  Immediately move away from the equipment or object. 

Electromagnetic  Equipment/  Environments

—Avoidance  of  high  electromagnetic  equipment  radiators  or 

environments is highly encouraged. Examples of equipment and/or environments include the following: 

  High-power  amateur  transmitters/antennas  or  citizen  band  (CB)  radio  or  Ham  radio  used  for  private 

recreation, communication, and wireless experimentation. 

  Electric arc welding or resistance welding equipment used for melting and joining metals or plastics. 

  Industrial  electric  induction  furnace/heater  or  electric  arc  furnace/heater  used  for  melting  metals  and 

plastics. 

  High-voltage areas identified by fenced areas, restricted access signs, and caution signs (safe if outside 

the fenced area). 

  Microwave transmitters identified by fenced areas, restricted access signs, and caution signs (safe if outside 

the fenced area). 

  Television, cell phone and radio towers identified by fenced areas, restricted access signs, and caution 

signs (safe if outside the fenced area). 

  Linear power amplifiers used for increasing the power output of radio transmitters, wireless communication 

applications, audio equipment or other electronic equipment. 

  Radio telemetry equipment used for tracking location of vehicles, equipment or animals. 

Use  of  accessories,  transducers  and  cables  other  than  those  specified  or  provided  by  the  manufacturer  of  this 
equipment  could  result  in  increased  electromagnetic  emissions  or  decreased  electromagnetic  immunity  of  this 
equipment and result in improper operation. 

Summary of Contents for Nalu Neurostimulation System

Page 1: ...Nalu Medical Inc Nalu User s Kit Instructions for Use PN MA 000007 Rev C NALU NEUROSTIMULATION SYSTEM USER INSTRUCTIONS FOR USE...

Page 2: ...urces ISO 15223 1 5 3 2 Keep dry A medical device that needs to be protected from moisture ISO 15223 1 5 3 4 Temperature limit The temperature limits to which the medical device can be safely exposed...

Page 3: ...nded for continuous operation IEC 60417 5333 Protected against access to certain hazardous parts Protected against solid objects over 12 5mm e g a finger and protected against falling drops of water i...

Page 4: ...ndications 7 Warnings 8 Precautions 11 Adverse Event Summary 13 Therapy Disc 16 Description 16 Therapy Disc Features 16 Powering ON the Therapy Disc 17 Connecting the Therapy Disc with the Nalu Implan...

Page 5: ...rger 25 Troubleshooting 25 Safety and Technical Checks 26 Cleaning Your Device 26 Cleaning the Therapy Disc 26 Cleaning the Trial Therapy Disc 26 Therapy Disc Charger and Clip Disposal 26 Environmenta...

Page 6: ...d medical professionals The Therapy Disc is designed to be worn when stimulation is needed Stimulation programs can be selected and adjusted from the Therapy Disc or a remote control application A Nal...

Page 7: ...rms and can be optionally used to control and manage Therapy Discs over a secure Bluetooth Low Energy connection Charger The Nalu charger recharges the batteries inside both the Therapy Discs SAFETY I...

Page 8: ...ury to the user In rare cases as a result of the unexpected changes in stimulation users have fallen down and been injured If you suspect that your Nalu Neurostimulation System is being affected by EM...

Page 9: ...he Nalu Neurostimulation System with other active implantable or body worn medical devices Malfunction and or damage could occur to either system that could result in harm to the user or other people...

Page 10: ...off or may cause permanent damage to the Nalu Implantable Pulse Generator particularly if used in close proximity to the device Lithotripsy Electrocautery Do not use monopolar cautery External defibri...

Page 11: ...elt may cause a fall or strangulation Tampering Do not modify or tamper with any component of the Nalu Neurostimulation System Tampering with the device could result in harm If the device is not worki...

Page 12: ...Therapy Disc may become more susceptible to damage by immersion Clean the Therapy Disc When needed clean the outside of the Therapy Disc with a damp cloth to prevent dust and dirt Storage Temperatures...

Page 13: ...location with or without user movement Persistent post surgical pain at hardware implantation sites Seroma or hematoma at surgery sites Spinal cord injury and or compression with subsequent neurologi...

Page 14: ...n hyperesthesia Change in stimulation that are possibly related to tissue changes around the electrodes shifts in electrode position loose electrical connections and lead or extension fractures which...

Page 15: ...Nalu User s Kit Instructions for Use PN MA 000007 Rev C Page 15 of 44...

Page 16: ...s for Use for details Battery life is optimal when the Therapy Disc is placed directly over the implant You are provided with two Therapy Discs in the initial Nalu User s Kit one to be used while the...

Page 17: ...e the Therapy Disc over the implant 2 Three 3 short beeps will sound when the Therapy Disc connects to the implant NOTE The Therapy Disc starts stimulation at the most recently used program and stimul...

Page 18: ...cted the green LED will blink four 4 times NOTE The Therapy Disc cycles through the programs in order When the last program is reached the next program will be Program 1 NOTE The maximum number of pro...

Page 19: ...Short Press on the Down Button 1 green blink per press Stimulation decreased 2 short orange blinks 2 short beeps Minimum stimulation level reached Long Press on the Down Button Dismiss reactivate ale...

Page 20: ...ions from the H100 Adhesive Clip switch to the H300 Adhesive Clip 34005 003 If the H100 Adhesive Clip is not sticky enough switch to the H200 Adhesive Clip 34005 001 for easier clip removals Preparing...

Page 21: ...9 Adhesive Wear Time 1 The wear time for the Adhesive Clip varies per activity level and skin type 2 Wear the clip until it feels like it is coming loose or starts to feel uncomfortable Removing the...

Page 22: ...nger and pull out the device as shown in Figure 13 Wearing the Relief Belt 1 Wear the belt over a thin layer of clothing at your waist Ensure that the belt is positioned so that the Therapy Disc butto...

Page 23: ...et location 1 Place the device over the implant location over a thin layer of clothing and wrap the strap around your limb 2 Place the strap through the loop and pull the strap to tighten the limb cuf...

Page 24: ...ring Up the Charger 1 Use only the provided AC DC power supply Plug the power supply in easily accessible socket so that it is easy to disconnect when necessary 2 Connect the USB C cable from the Powe...

Page 25: ...The stimulation amplitude is set too low Increase the amplitude The Therapy Disc is Unresponsive The Therapy Disc is not powered ON Turn the power ON The Therapy Disc is Unresponsive The Therapy Disc...

Page 26: ...odic safety checks or maintenance of the Therapy Disc are not required The Therapy Disc contains no user serviceable parts If repair or service is needed contact your clinician or a Nalu representativ...

Page 27: ...evices Carry this card with you at all times If you move change doctors or lose your card contact Nalu for a replacement card Refer to the contact information at the end of this manual WHEN TO CALL YO...

Page 28: ...stimulation ID Card 1 MA 000007 Nalu Neurostimulation User Instructions for Use 1 53002 001 Remote App 74002 Nalu Neurostimulation Trial User s Kit 34002 Trial Therapy Disc 2 34007 Relief Belt 1 43001...

Page 29: ...Weight 100g Trial Therapy Disc 34002 Description Case Material Polycarbonate Acrylonitrile Butadiene Styrene Diameter 3 76 mm Thickness 0 6 15 mm Weight 100g Description Cable Material Polyvinyl Chlo...

Page 30: ...9 g Adhesive Clip Pack H300 34005 003 Adhesive Clip Pack H100 34005 002 Description Clip Description Clip Material Polypropylene Material Polypropylene Diameter 3 76 mm Diameter 3 76 mm Description Ad...

Page 31: ...ption Velcro Material Neoprene Material Acetal Length 18 46 cm Weight 33 g Limb Cuff Large 34011 Description Cuff Description Velcro Material Neoprene Material Acetal Length 25 64 cm Weight 40 g Thera...

Page 32: ...FCC ID 2AAQS ISP1507 Trail Therapy Disc Contains FCC ID 2AAQS ISP1507 Classification per IEC 60601 1 Therapy Disc and Trial Therapy Disc In use Type BF Charging Class II Quality of Service for Wireles...

Page 33: ...le and cordless telephones and tablets can affect the device Statement of FCC Compliance This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to par...

Page 34: ...ity Test IEC 60601 Test Level Compliance Electromagnetic Environment Guidance Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 V Portable and mobile RF communications equipment should be used no...

Page 35: ...Generators Dual Integrated Model 11002 040 Single Integrated Model 11001 040 Nalu Implantable Pulse Generators Dual Ported Model 11004 002 connected to two 40 cm Nalu Lead Model 12001 040 Single Porte...

Page 36: ...ve RF coil that includes a knee foot ankle or wrist transmit receive RF coil No part of the implanted Nalu Neurostimulation System may be within one of these transmit receive RF coils Under the scan c...

Page 37: ...ive transmit receive RF coil Head Brain MRI Examination Figure 1 Head brain MRI examinations are permitted using a 1 5 T or 3 T MRI system and a transmit receive RF head coil No part of the implanted...

Page 38: ...ser s Kit Instructions for Use PN MA 000007 Rev C Page 38 of 44 Extremity MRI Examinations Patient implanted with Nalu Neurostimulation System Figure 2a 1 5 T or 3 T MR system Transmit receive RF knee...

Page 39: ...Nalu User s Kit Instructions for Use PN MA 000007 Rev C Page 39 of 44 Transmit receive RF foot ankle coil 1 5 T or 3 T MR system Patient Implanted with Nalu Neurostimulation System Figure 2b...

Page 40: ...facturer attached to the Nalu Implantable Pulse Generator The risk of performing an MRI examination under those circumstances has not been evaluated and thus may cause harm to the patient and or the c...

Page 41: ...impedance check Do not perform an MRI if the impedance is greater than 10 k Remove the Therapy Disc from the patient before entering the MRI system room Do not conduct an MRI examination if the 40 cm...

Page 42: ...coil at 1 5 T or 3 T For head brain MRI examinations only the transmit receive RF head coil is permitted for use No parts of the implanted Nalu Neurostimulation System may be within the transmit recei...

Page 43: ...ulation System may be within the transmit receive RF head coil Preparation of the Patient Prior to the MRI Examination Inform the patient of the risks associated with undergoing an MRI examination an...

Page 44: ...ort nalumed com www nalumed com This product can expose you to chemicals including ethylene oxide which is known to the State of California to cause cancer and birth defect or other reproductive harm...

Reviews: