![Nalu Medical Nalu Neurostimulation System User Instructions For Use Download Page 37](http://html1.mh-extra.com/html/nalu-medical/nalu-neurostimulation-system/nalu-neurostimulation-system_user-instructions-for-use_3149850037.webp)
Nalu User’s Kit Instructions for Use PN: MA-000007 Rev C
Page
37
of
44
Acceptable 1.5 T/64-MHz or 3 T/128 MHz MRI Scenarios
Note the position of the implantable pulse generator (IPG) and leads relative to the transmitted RF energy
for each respective transmit/receive RF coil.
Head/Brain MRI Examination
Figure 1
. Head/brain MRI examinations are permitted using a 1.5 T or 3 T MRI system and a transmit/receive RF
head coil. No part of the implanted Nalu Neurostimulation System may be within the transmit/receive RF head coil.
All other aforementioned conditions must be carefully followed.
Transmit / receive
RF head coil
Patient implanted with Nalu
Neurostimulation System
1.5 T or 3 T MR system