Adhesive Keeps Falling Off
Improper wear period.
Reduce your wear time. Replace the adhesive
clip before it falls off.
Talk to your Nalu representative about
switching to a different adhesive.
Adhesive Causing Skin Breakage
The adhesive was removed too
early.
Increase your wear time and remove the
adhesive when it is looser.
Adjust the next clip location to allow the
damaged skin to heal.
Use the relief belt for a period of time to allow
the skin to heal.
Talk to your Nalu representative about
switching to a different adhesive.
Adhesive Causing Excessive Skin
Irritation, Including Increased
Redness, Itchiness, and
Bumpiness
Possible allergic reaction.
Contact your clinician.
SAFETY AND TECHNICAL CHECKS
Periodic safety checks or maintenance of the Therapy Disc are not required. The Therapy Disc contains no user-
serviceable parts. If repair or service is needed, contact your clinician or a Nalu representative for a replacement.
Refer to the contact information at the end of this manual.
CLEANING YOUR DEVICE
Cleaning the Therapy Disc
When needed, clean the outside of the Therapy Disc with a damp cloth to prevent dirt and dust. Do not use abrasive
or caustic cleaning products on your Therapy Disc.
Cleaning the Trial Therapy Disc
Clean the outside of the Trial Therapy Disc with a damp cloth when needed to prevent dirt and dust. Do
not
use
abrasive or caustic cleaning products on your Trial Therapy Disc.
THERAPY DISC, CHARGER, AND CLIP DISPOSAL
The Therapy Disc and charger should be returned to your clinician or a Nalu representative. Do not dispose of your
Therapy Disc or charger in the garbage.
For the single-use adhesive clips, adhere to local disposal requirements / regulations.