![Nalu Medical Nalu Neurostimulation System User Instructions For Use Download Page 20](http://html1.mh-extra.com/html/nalu-medical/nalu-neurostimulation-system/nalu-neurostimulation-system_user-instructions-for-use_3149850020.webp)
Nalu User’s Kit Instructions for Use PN: MA-000007 Rev C
Page
20
of
44
ADHESIVE CLIP
Figure 2: Clip
Figure 3: Clip (Back)
Figure 4: Clip and Therapy Disc
Description
The single-use Adhesive Clip is designed to hold the Therapy Disc on the body using a skin friendly adhesive. The
Adhesive Clip is placed over the implant and the Therapy Disc is inserted into the clip. The Adhesive Clip is not
intended to be reused.
Your user kit contains the H100 Adhesive Clips (34005-002). If you experience moderate to severe skin reactions
from the H100 Adhesive Clip, switch to the H300 Adhesive Clip (34005-003). If the H100 Adhesive Clip is not sticky
enough, switch to the H200 Adhesive Clip (34005-001) for easier clip removals.
Preparing Your Skin
1. Make sure your skin is clean and dry.
2. If necessary, trim hair with scissors or small beard trimmer.
Do not shave
since this can lead to irritation.
3. If desired, commercially available skin barrier wipes and sprays may be used to prep the skin.
Applying the Clip
1. Remove the adhesive liner, as shown in Figure 5.
2. Grasp the clip between your thumb and finger and place your index
finger through the center hole. The clip should be held, as shown in
Figure 6.
Figure 5 Remove Liner
Figure 6 Clip Orientation