Nahita LNB003 Instruction Manual Download Page 9

 

 

nP- 

Maximum 

power 

output 

The  maximum  power  percentage  of  the  heating 
output. 

(0

100%) 100 

Co- 

Turn off the 

heating 

output 

deviation 

When  the  "temperature  measurement  value 

 

temperature  set  value  +  Co",  the  heating  output  is 
turned off. 

(0.0

20.0

5.0 

SPL 

Lower 

temperature 

setting 

Lower limit of temperature set point 

(-50.0

100.0

0.0 

SPH 

Upper 

temperature 

limit 

Upper limit of temperature set point 

(SPL

100.0

60.0 

Adr 

mailing 

address 

The local communication address. 

(1

16) 1 

 
 

Internal parameter table -3

 

Parameter 

indication

 

Parameter 

name

 

Parameter function description

 

Factory value

 

Lc

 

 

Password

 

The  ambient  temperature  can  be  viewed  when 

"Lc=18". 

 

Ht 

Ambient 

temperature 

The ambient temperature at which the controller is 
located. 

 

 

2.

 

Speed internal parameters setting 

 
Press  and  hold  the 

Speed

  button  for  about  3  seconds.  The  temperature  display  area 

displays the password prompt 

Lc

. The speed display area displays the password value. You 

can change the password value to the required password value by adding the key, decreasing 
key  and  shift  key.  Then click  the 

Speed

  button.  If  the  password  value  is  incorrect,  the 

controller  will  automatically  return  to  the  normal  display  state.  If  the  password  value  is 
correct, it will enter the speed internal parameter setting state, and then click the 

Speed

 

button to modify each parameter in turn. After the modification, press and hold the 

Speed

 

button for 3 seconds to exit the speed internal parameter setting state, and the parameter value 
is automatically saved. 

 

Note: The speed parameter is forbidden to be modified during the running of the controller. If it needs 

to be modified, please stop the controller and then modify it. 

 
 
Speed---internal parameter table (1) 
 

Parameter   

indication

 

Parameter 

name

 

Parameter function description

 

(Range) Factory Value

 

Lc 

Password

 

The  parameter  value  can  be  viewed  and 

modified when "Lc=3".

 

Pd- 

Proportional  Speed proportional gain.

 

(1

100) 10 

Summary of Contents for LNB003

Page 1: ...Version 1 July 2021 Instruction Manual NAHITA BLUE SHAKING INCUBATOR REFERENCE LNB003...

Page 2: ...r the appearance of the product is intact and the random accessories matches the packing list If the product model you received is inconsistent with your booking or the appearance is damaged or there...

Page 3: ...et hands otherwise there is a risk of electric shock 4 Before any repair or maintenance is carried out the power must be disconnected to prevent electric shock or injury 5 Please wear gloves when repa...

Page 4: ...als such as gasoline or thinner Keep the inside and outside of the box clean often cleaning debris and smudges 1 The large screen LCD displays the temperature speed and working time continuously accur...

Page 5: ...W Controller Temp Control Mode LCD PID intelligent control Cyclotron Control Mode Automatic changes of rotation direction Setting Mode Touch button setting Temp Display Mode LCD display Speed Display...

Page 6: ...key thermometer symbol press enter temperature parameter setting state then click return to normal interface Speed function key with arrowhead circle press enter speed parameter setting state then pr...

Page 7: ...0 5 0 Ct Compressor start delay Compressor start delay protection time the minimum time interval from compressor stop to restart 0 600s 180 uP compressor Start threshold When the compressor is workin...

Page 8: ...essor 2 compressor according to the value of Htd see below to determine the balanced or intermittent work of the compressor 0 2 0 CP Compressor working mode fixed switching point point When ndc 1 If t...

Page 9: ...password prompt Lc The speed display area displays the password value You can change the password value to the required password value by adding the key decreasing key and shift key Then click the Sp...

Page 10: ...tion dIF 0 Specifies that the clockwise direction of rotation is positive dIF 1 Specifies that the counterclockwise direction of rotation is positive 0 1 0 FdS Speed feedback value Speed feedback coef...

Page 11: ...by shifting reducing increasing etc then click the Time button to enter the stop timing time the right display prompt p setting status Also modify to the desired setting value then click the Time butt...

Page 12: ...light up and enters after about 3 seconds Normal display status 2 Temperature Press the Temperature button the number on the right side of the temperature display area flashes use the shift button de...

Page 13: ...the operation Notes 1 If the set value of the timing time is changed during the running the controller restarts the timing according to the new set time 2 The timing mode can be selected by modifying...

Page 14: ...ng state the controller will automatically return to the normal display state 5 If the display window of the controller shows to indicate the fault of the temperature sensor or the controller itself p...

Page 15: ...ed smoothly Adjust the foot of the instrument to make the instrument smooth Foreign body barrier at the bottom of the plate Remove foreign body Control circuit fault Notify distributor to repair The m...

Reviews: