non-compliance with the following safety precautions happens, the centrifuges would
cause the operator or other personnel injury or make damages to centrifuge separation
plant and the internal sample:
●
Centrifuge design is neither anti-corrosion, nor is proof, therefore, ensure no
centrifuges corrosion in the environment and the possibility of an explosion happen in
the use of the environment;
●
The use of centrifuges is strictly prohibited in the following materials:
- Flammable and explosive materials;
- Strengthen the role of materials science;
- Toxic or radioactive substances or pathogenic micro-organisms and so on.
●
For the separation of corrosive substances and pathogenic micro-organisms cell
effective sealing measures should be carried out in advance and effective disinfection
measures should be carried out after use. Details, see “repair and maintenance issues -
sterilization” in the contents.
●
The substances for the separation of corrosive can cause damage and destruction of
centrifuge internal material or weaken the rotor mechanical strength, so when
separate of corrosive substances , the corrosive substances must be kept in the
protective container.
1.1
Operation Notes
●
Before centrifuge run (sample separation), it must be confirmed with appropriate
rotor and ensure a solid installation.
●
When in the process of running centrifuges ( rotors turning) or in the process to
stop centrifuge ( but when the rotor is still rotating), do not manually open the door or
move the centrifuge;
●
The components used in Magal low-speed Benchtop Air-refrigerate Centrifuge
must be dedicated accessories provided byMagal.. A number of common components,
such as: glass and plastic containers used for the separation, it should be confirmed to
meet the requirements to use the products, which shall conform to the corresponding
rotor speed and the largest of the greatest use of centrifugal force, etc.
●
Do not open the door of the cases when using the centrifuges or in separation of
sample;
●
The replacement of the mechanical parts of the centrifuge and electronic devices,
must be implemented by relevant personnel designated by theMagal.;
●
When the operator uses the centrifuges, it is important to choose an appropriate
load of the rotor, and shall not overload the use of the rotor;
●
Ensure regular checks of the rotor, if the rotor has obvious corrosion or obvious
signs of damage, using must be stoped;
●
After using for a period of time, the provisions of maintenance should be strictly in
accordance with the "cleansing and disinfection,"
1.2
The relevant quoting standards
●
GB4793.1-2007 Part I of Safety Requirements of Measurement, control of