![Nahita GLC001 Operation And Maintenance Manual Download Page 25](http://html1.mh-extra.com/html/nahita/glc001/glc001_operation-and-maintenance-manual_4528703025.webp)
English
User's Manual
of GLC001
Version 1, July 2023
- 24 -
●
Turn on the power switch and wait for the self check to complete;
●
Turn on the power switch and wait for the self check to complete; Press the
key to open the centrifuge door cover;
●
Using the special tool provided with the machine for disassembling and
disassembling the rotor, loosen the locking sleeve and take out the rotor. Clean the
locking sleeve. Note that the clockwise direction is for tightening the locking
sleeve, and the counterclockwise direction is for loosening the locking sleeve;
●
Clean the conical surface of the motor spindle without leaving any dirt. Add an
appropriate amount of lubricating oil or apply it with lubricating paper;
6.1.3.8
When disassembling the machine, the power must be cut off first and the
power cord connected to the back wall of the machine must be unplugged. It is not
allowed to operate with electricity to prevent personnel from getting electric shock or
damaging the machine. Attention: This operation can only be carried out by our
company's maintenance personnel who have received specialized training!
6.1.3.9
This machine can only use parts provided by our company.
6.1.3.10
When the centrifuge is not in use, the power should be cut off.
6.1.3.11
Transportation and storage
This machine is a precision instrument and equipment. During
transportation and storage, please pay attention to moisture and shock resistance, and
do not lay it horizontally or upside down.