![Nahita GLC001 Operation And Maintenance Manual Download Page 21](http://html1.mh-extra.com/html/nahita/glc001/glc001_operation-and-maintenance-manual_4528703021.webp)
English
User's Manual
of GLC001
Version 1, July 2023
- 20 -
Figure 7: Example parameter setting completed
5.10 Parameter settings Other
5.10.1
Relative centrifugal force setting:
Press the
key twice
continuously on the control panel, causing the number at Rcfto flash
→
turn the parameter adjustment knob key to set its value. Note: The Speed
value is automatically converted with the Rcm value;
5.10.2
During the parameter setting process, if an alarm occurs due to a machine
malfunction or incorrect parameter settings, press the
key to cancel the alarm
and then refer to the rotor number again for setting.
5.10.3
Press the
key and the machine will start running (if you
need to stop halfway, please press the
key). The set time will
gradually decrease from the set value to zero. When the time value shows
zero, the centrifuge will automatically stop, and the speed will gradually
decrease from the set value to zero (the time required for the speed to
decrease from the set value to zero is related to the speed reduction
setting). When the speed becomes zero, the door cover will automatically
open and the machine will emit shutdown music, Press the
key to
stop the sound. Centrifuge completion.
5.10.4
If instantaneous centrifugation is required:
Please press and hold
the
key on the control panel, and the speed will continue to
increase. Release the key to stop, and the maximum speed during this time
will be the set speed corresponding to the rotor number and increase in
Sec. unit time.
5.10.5
Reminder tone settings:
Press the
key to adjust the prompt
volume.
5.10.6
After the machine speed stabilizes, if necessary, the parameters such as
speed/centrifugal force, time, and lifting speed can be modified again. After resetting