Page 11 of 19
Pairing Accessories
The MXD-LTE is capable of pairing a large number of accessories or devices to one of many available
zones. Each of these accessories can be paired in the same zone or individual zones. For more
information on available zones please contact your service provider.
Note: information about available
When an alarm call is made to the monitoring center, the MXD-LTE provides them with the device type
and the zone the device is associated with. Zone is often a generic way to associate a device to a specific
user or area of a house. Many users will pair several devices of the same type (such as PHBs) to the
same zone. However, the monitoring center cannot distinguish which specific device was activated. The
MXD-LTE may also have similar types of devices paired into separate zones, allowing the monitoring
center to specifically identify which device was activated.
For example, if two individual’s PHBs are programmed to the same zone, the monitoring center will
know that a PHB was pressed, but not which one. If they are paired to different zones, the monitoring
center can distinguish which PHB was pressed.
Similarly, if two Smoke Detectors are paired into zone 1, the monitoring center will know a Smoke
Detector was activated, but not which one. However, if the bedroom Smoke Detector is paired to zone 1
and the kitchen Smoke Detector is paired to zone 2, the monitoring center will know which Smoke
Detector was activated.
Different device types may be paired to the same zone or to different zones and the monitoring center
will know which was activated so long as there are not multiple devices of the same type paired to the
same zone.
The following is a list of typical device types the MXD-LTE is compatible with:
•
Personal Help Buttons, Wall Mount Buttons and Fall Detectors (PHBs)
•
Medication Dispensers
•
Smoke Detectors
•
CO Detectors
•
Voice Activation Devices
Please note, Fall Detectors are paired as a PHB devices. When activated by a button press, a PHB alarm
is transmitted to the monitoring center for the associated zone. If a fall is detected, a fall alarm is
transmitted to the monitoring center for the same zone as if it were a PHB.
The pairing of devices is usually performed by your Dealer before you receive your MXD-LTE. The
following is provided for reference. If you have questions, please contact your dealer.
NOTE: IT IS IMPERATIVE THAT THE MONITORING CENTER IS MADE AWARE OF EVERY DEVICE AND THE
CORRESPONDING ZONE THAT YOUR MXD-LTE HAS PAIRED.