5.5.
Jumper-settings
Jumper JP1
– JP3:
Via Jumper JP3 the analog output is switched from constant current (mA) to voltage (V).
At the same time, the Jumper JP1 and JP2 have to be reconnected, thereby it is possible
to measure the selected output signal at the test socket in the front.
Jumper JP5, JP6, JP7, JP8 and JP9:
Measuring input for:
JP-5
JP-6
JP-7
JP-8
JP-9
Voltage (V)
X
X
X*
X
X
Voltage (mV)
X
--
X*
X
X
Current (mA)
X
X
X*
X
X
Thermocouple
with internal TVG
X
--
X*
X
X
Thermocouple
with external TVG
X
--
--
X
X
Thermocouple
with thermostat
X
--
X**
X
X
Pt-100/2-
wire-circuit
X
X
--
X
X
Pt-100/3-
wire-circuit
X
X
--
X
--
Pt-100/4-
wire-circuit
X
X
--
--
--
Potentiometer/0-600
2-wire-circuit
X
X
--
X
X
Potentiometer/0-600
3- wire- circuit
X
X
--
X
--
Potentiometer/0-600
4- wire- circuit
X
X
--
--
--
Potentiometer/0-5000
2- wire- circuit
--
X
--
X
X
Potentiometer/0-5000
3- wire- circuit
--
X
--
X
--
Potentiometer/0-5000
4- wire- circuit
--
X
--
--
--
X* = JP7 in Position Pt100, X** = JP7 in Position Thermostat
X = Jumper closed
-- = Jumper opened
After pressing the button
JUMPER FINDER
in the WINSMART’s masks
Measuring
Input
and
Analog Output
, the position of the jumpers is shown according to the
component diagram section.