Running the Example HCF Test
Landmark™ Test System Operation
103
A. From the Station Manager application main window, click the Manual Command
button to open the Manual Command window.
B. Select the following parameters:
l
Channel
— Landmark 1 (or other depending on your configuration)
l
Control Mode
— Force
l
Enable Manual Command
— Checked
C. Type
0
in the
Manual Cmd
text box and press
Enter
.
3. Switch to Displacement control mode.
This prevents unwanted movement of the actuator.
A. In the Station Manager application main window, click the
Manual Control
button.
B. In the Manual Controls window Control Mode drop-down list, select
Displacement
.
4. Release the upper grip.
Use the controls on the load frame.
5.
Warning:
Do not move the control slider until the crush zone is clear, and you can monitor
actuator movement.
Unskilled movement of the actuator can result in machine damage or personal injury.