EVOLUTION 6000 Thermal Imaging Camera
40
Maintenance
MSA
US
5.3
Germanium Lens Replacement
The camera is designed for simple replacement of a damaged Germanium lens.
Fig. 16
Germanium lens retainer, parts and tools
Fig. 17
Germanium lens retainer, screws to be removed
To replace the germanium lens:
(1)
Remove the battery from the camera.
(2)
Using a 2.5 mm Allen wrench, remove the six screws holding the germanium lens retainer.
(3)
Remove the broken germanium lens and any dirt that may have entered the sealing area.
(4)
Insert the new lens in the retainer and reinstall on the camera.
(5)
Tighten the six screws to 1.6 Nm (14 in-lbs.) of torque.
WARNING
The germanium lens must be replaced immediately after damage is detected. Failure to do so
may allow water and other contaminants to enter the camera body, resulting in a possible loss
of camera function and severe damage to electronic components. Should water or other contam-
inants enter the camera housing, the camera must be returned to an authorized MSA repair fa-
cility. Do not operate the camera with a damaged germanium lens.