Camera Setup
EVOLUTION 6000 Thermal Imaging Camera
35
US
4.3
Options
Selecting Options
NOTE: Options not installed on the camera do not display.
NOTE: Options locked by the MSA FireService Utility will appear grayed out and cannot be se-
lected.
Configuring Options
For all options, the current selection displays
Button
Functions as
ZOOM button
SCROLL button (scrolls through available options)
PALETTE button
SELECT button (selects the option)
Option
Description
Exit
To exit configuration menu
and return to normal opera-
tion
Spotmeter
(Digital
Tempera-
ture Target)
Temperature Display Set-
ting-°F or °C selection for the
spot meter display
(1)
SCROLL to either °F or °C and SELECT to change the
current selection.
Selection is confirmed and camera returns to the
main Config menu.
Compass
Compass ON/OFF,
Display Config and Calibra-
tion (if local control was not
disabled via MSA FireServ-
ice Utility package)
(1)
SCROLL to COMPASS ON or COMPASS OFF and
SELECT to turn compass ON or OFF.
(2)
SCROLL to SET DISPLAY TYPE TEXT and SELECT
to select text icons (displayed as N, S, E, W, NW, etc).
(3)
SCROLL to SET DISPLAY TYPE ICON and SELECT
to select a compass icon to indicate direction.
(4)
SCROLL to CALIBRATE and SELECT to calibrate the
compass.
It is recommended that the compass be recalibrat-
ed at least monthly, especially if camera is subject-
ed to strong magnetic fields.
(5)
Choose YES to proceed with the calibration and NO to
exit calibration.
(6)
Follow the on-screen instructions.
(7)
Hold the camera in front of you and face north.
(8)
Tumble the camera in all three axes.
When ALL DONE displays, the compass is suc-
cessfully recalibrated.
If "Calibration Failed" displays, exit and try again.
(9)
SELECT EXIT to exit compass calibration.
Selection is confirmed and camera returns to the
Config main menu.