Operation Manual
9 of 18
ENVIRONMENT:
Ambient Operating Temperature Range : 15-40˚C
Storage Temperature Range : 10-50˚C
Humidity range : 40-75% RH
READOUT SPECIFICATIONS:
Resolution : 1˚C / 1˚F
Temperature units : ˚C/˚F
POWER SUPPLY:
Line Voltage / Frequency : 110 VAC ±10% / 60Hz.
Power consumption : 1850VA
OPTIONAL RS232 COMMUNICATION INTERFACE:
Type of connection : 9 pole, D sub male
Software : Calibration software
OPTIONAL FORCED COOLING:
Air cooling : Forced Air cooling (Dry Compressed Air required)
THERMAL SPECIFICATIONS:
Maximum Temperature : 1200˚C / 2192˚F
Minimum Temperature : 300˚C / 572˚F
Basic Accuracy: Better than ±3˚C.
Stability : ±0.5˚C up to 600˚C & ±1˚C above 600˚C
Heating time to max. (incl. Insert) : 60 minutes
Cooling time (without Forced Air Cooling) : 2 ½ hours
Cooling time (with forced air cooling): 45 min. from 1200˚C to 300˚C.
Multi hole insert : (4*7mm) + (3*9mm) or to suit