communications equipment (transmitters) and the
eMOTION
as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d =1,2
√P
80 MHz to 800 MHz
d =1,2
√P
800 MHz to 2,7 GHz
d =2,3
√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Guidance and manufacturer’s declaration – electromagnetic emissions
The eMOTION is intended for use in the electromagnetic environment (for home healthcare and professional
healthcare) specified below.
The customer or the user of the eMOTION should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment-guidance
(for home healthcare and professional healthcare
environment)
RF emissions CISPR 11
Group 1
The
eMOTION uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The eMOTION is suitable for use in all establishments,
including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Compliance