Manufacturer’s declaration-electromagnetic immunity

 

The 

eMOTION

  is  intended  for  use  in  the  electromagnetic  environment  (for  home  healthcare  and  professional 

healthcare) specified below. 
The customer or the user of the 

eMOTION

 should assure that it is used in such an environment.

 

Immunity test 

IEC 60601 

test level 

Compliance level 

Electromagnetic environment-guidance 

(for home healthcare and professional 

healthcare environment) 

Electrostatic 
discharge 
(ESD) IEC 
61000-4-2 

Contact:±8 kV 

Air±2 kV,±4 
kV,±8 kV,±15 kV 

Contact:±8 kV 

Air±2 kV,±4 
kV,±8 kV,±15 kV 

Floors should be wood, concrete or ceramic tile. If 
floors are covered with synthetic material, the 
relative humidity should be at least 30% 

Electrical fast 
transient/burst IEC 
61000-4-4 

+ 2kV for power
supply lines

+ 1kV for
input/output lines

+ 2kV for power
supply lines

Not applicable 

Mains power quality should be that of a typical 
home healthcare and professional healthcare 
environment. 

Surge IEC 
61000-4-5 

+ 0.5kV, +1kV
line(s) to line(s)

+ 0.5kV, +1kV,+
2kV line(s) to
earth

+ 0.5kV, +1kV
line(s) to line(s)
Not applicable

Mains power quality should be that of a typical 
home healthcare and professional healthcare 
environment. 

Voltage Dips, short 
interruptions and 
voltage variations 
on power supply 
input lines IEC 
61000-4-11 

Voltage dips: 

0 % 

U

T; 0,5 cycle 

0 % 

U

T; 1 cycle 

70 % 

U

T; 25/30 

cycles 

Voltage 
interruptions: 

0 % 

U

T; 250/300 

cycle 

Voltage dips: 

0 % 

U

T; 0,5 cycle 

0 % 

U

T; 1 cycle 

70 % 

U

T; 25/30 

cycles 

Voltage 
interruptions: 

0 % 

U

T; 250/300 

cycle 

Mains power quality should be that of a typical 
home healthcare and professional healthcare 
environment. If the user of the eMOTION requires 
continued operation during power mains 
interruptions, it is recommended that the 
eMOTION be powered from an uninterruptible 
power supply or a battery. 

Power frequency 
(50, 60 Hz) 
magnetic field IEC 
61000-4-8 

30 A/m 

50 Hz or 60 Hz 

30 A/m 

50 Hz, 60 Hz 

The

 

eMOTION power frequency magnetic fields 

should be at levels characteristic of a typical 
location in a typical home healthcare and 
professional healthcare environment. 

NOTE 

UT is the a.c. mains voltage prior to application of the test level. 

* During DIP interference, the pump will outage these are normal. The pump outage does not affect the motor

operation.

Summary of Contents for eMOTION

Page 1: ...for future reference eMOTION controller User Manual Part Number FC MOX1006 controller and FM MOX1006 mattress Manufactured by MOXI ENTERPRISES LLC 1939 S VANDEVENTER AVE ST LOUIS MO 63110 Telephone 31...

Page 2: ...vent from personal injury or damage to the system Moxi Enterprise LLC assumes no responsibility for any damage or injury caused by improper assembly or use of this product WARNING Always unplug the co...

Page 3: ...ho are at least fifteen years in age with the ability to read and understand English and Westernized Arabic Numerals This device should not be operated by patient 3 FEATURES The eMOTION Alternating Pr...

Page 4: ...n is turned on it automatically increases pressure level to support patient when needed especially during head raised position and during ingress egress movements Lockout function The pump is also equ...

Page 5: ...face for tears or cracking Do not use if any damage presents Notify your authorized distributor of the damages 4 Connect pump hoses to mattress connect the main connection tube of the mattress to the...

Page 6: ...roper care and maintenance are essential to keeping eMOTION Alternating Pressure Mattress System in a safe operating condition In addition to inspecting the unit before each use periodic maintenance c...

Page 7: ...ssly soiled the cover should be removed cleaned and decontaminated 7 Using a clean sponge or paper towel wipe down the surface and cells with a diluted detergent solution or cleaner disinfectant or ot...

Page 8: ...e a power regulator Faulty fuse Replace the fuse Patient is bottoming out Inadequate pressure level for the patient Adjust comfort setting to o1 or 2 levels higher and wait for a few minutes for best...

Page 9: ...100 240V 50 60 Hz 0 2 0 1A Fuse rating T2AL 250V Classification Electrical Class II Type BF Not AP or AGP type Temperature Operation 41 F 104 F 5 C 40 C Storage Transport 41 F 140 F 5 C 60 C RH Relati...

Page 10: ...efects in materials and workmanship 1 year from the date of purchase for the original consumer purchaser This device was built to exacting standards and carefully inspected prior to shipment We offer...

Page 11: ...Electromagnetic propagation is affected by absorption and reflection from structures objects and people Guidance and manufacturer s declaration electromagnetic emissions The eMOTION is intended for u...

Page 12: ...to earth 0 5kV 1kV line s to line s Not applicable Mains power quality should be that of a typical home healthcare and professional healthcare environment Voltage Dips short interruptions and voltage...

Page 13: ...e transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnet...

Page 14: ...0 3 28 28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217 Hz 2 0 3 28 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7...

Page 15: ...FM MOX1006 mattress Manufactured by MOXI ENTERPRISES LLC 1939 S VANDEVENTER AVE ST LOUIS MO 63110 T l phone 314 771 5222 T l copieur 314 771 5225 www moxiusa com Veuillez lire le manuel avant de faire...

Page 16: ...ement NE PAS UTILISER LE PRODUIT Contacter le Soutien technique ou votre distributeur autoris pour viter des blessures personnelles ou des dommages au syst me Moxi Enterprise LLC n assume aucune respo...

Page 17: ...faire de fa on s curitaire en cas d urgence Lorsque les c t s de lit sont utilis s adh rer au d gagement minimum sugg r Si le d gagement minimum ne peut pas tre respect utiliser une rehausse pour le...

Page 18: ...ajuster le r glage de confort afin d atteindre le confort optimal du patient Ce mode devrait rester pas plus d une heure Fonction Max Lors du d marrage initial utiliser la fonction max pour permettre...

Page 19: ...patient Le Syst me de matelas pression alternative eMOTION a une capacit de poids maximal de 200 kg 440 lb 4 INSTALLATION Vue arri re pompe Vue avant pompe 6 Retirer l unit de contr le de la bo te et...

Page 20: ...CTIONNEMENT Illustration du Panneau de contr le 5 Appuyer et tenir le bouton POWER sur le panneau avant pendant une seconde le DEL POWER deviendra alors vert indiquant que la pompe fonctionne 6 La DEL...

Page 21: ...OTION peut changer la structure du Syst me de matelas et pourrait cr er une condition dangereuse ce qui pourrait causer des blessures s rieuses et qui annulera la garantie La pompe contient des compos...

Page 22: ...chaude et un d tergent neutre ou avec une solution hypochlorite de sodium eau de Javel 19 Ou sinon enlever la housse et laver 40 C 104 F en utilisant un d tergent normal Il est essentiel que les artic...

Page 23: ...er le fusible Patient est enfonc Niveau de pression inad quat pour le patient Ajuster le r glage de confort de 1 ou 2 niveaux de plus et attendre quelques minutes pour le meilleur confort L air sort d...

Page 24: ...15 20 minutes D bit 5 L min compresseur Pression 15 45 6 mmHg 7 niveaux de pression Tension d entr e nominale AC 100 240V 50 60 Hz 0 2 0 1A Calibre du fusible T2AL 250V Classification lectrique Class...

Page 25: ...avec la directive WEEE Veuillez tre responsable de l environnement et recycler ce produit travers votre installation de recyclage la fin de sa dur e de vie ou le jeter en accord avec les r glementatio...

Page 26: ...l metteur REMARQUE1 80 MHz et 800 MHz la distance d loignement pour la plage de fr quence plus lev e s applique REMARQUE2 Ces directives peuvent ne pas s appliquer toutes les situations La propagatio...

Page 27: ...ne s Non applicable La qualit de l alimentation r seau devrait tre celle d un environnement typique de soins domicile et de soins de sant professionnels Chutes de tension courtes interruptions et vari...

Page 28: ...tteur et d est la distance de s paration recommand e en m tres m Les intensit s de champs des metteurs RF fixes tel que d termin par un sondage de site lectromagn tique a devraient tre moins que le ni...

Page 29: ...28 870 930 1720 1700 1990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Modulation par impulsion b 217 Hz 2 0 3 28 28 1845 1970 2450 2400 2570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band...

Page 30: ...AL300446 V1 00...

Reviews: