11
Manufacturer’s declaration-electromagnetic immunity
The
eMOTION
is intended for use in the electromagnetic environment (for home healthcare and professional
healthcare) specified below.
The customer or the user of the
eMOTION
should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment-guidance
(for home healthcare and professional
healthcare environment)
Electrostatic
discharge
(ESD) IEC
61000-4-2
Contact:±8 kV
Air±2 kV,±4
kV,±8 kV,±15 kV
Contact:±8 kV
Air±2 kV,±4
kV,±8 kV,±15 kV
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%
Electrical fast
transient/burst IEC
61000-4-4
+ 2kV for power
supply lines
+ 1kV for
input/output lines
+ 2kV for power
supply lines
Not applicable
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment.
Surge IEC
61000-4-5
+ 0.5kV, +1kV
line(s) to line(s)
+ 0.5kV, +1kV,+
2kV line(s) to
earth
+ 0.5kV, +1kV
line(s) to line(s)
Not applicable
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment.
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
Voltage dips:
0 %
U
T; 0,5 cycle
0 %
U
T; 1 cycle
70 %
U
T; 25/30
cycles
Voltage
interruptions:
0 %
U
T; 250/300
cycle
Voltage dips:
0 %
U
T; 0,5 cycle
0 %
U
T; 1 cycle
70 %
U
T; 25/30
cycles
Voltage
interruptions:
0 %
U
T; 250/300
cycle
Mains power quality should be that of a typical
home healthcare and professional healthcare
environment. If the user of the eMOTION requires
continued operation during power mains
interruptions, it is recommended that the
eMOTION be powered from an uninterruptible
power supply or a battery.
Power frequency
(50, 60 Hz)
magnetic field IEC
61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz, 60 Hz
The
eMOTION power frequency magnetic fields
should be at levels characteristic of a typical
location in a typical home healthcare and
professional healthcare environment.
NOTE
UT is the a.c. mains voltage prior to application of the test level.
* During DIP interference, the pump will outage these are normal. The pump outage does not affect the motor
operation.