6
When you believe a component or part is not functioning normally, please contact MOXI
immediately as a potentially dangerous condition could exist.
Replacing air filters:
1. Switch off the electrical supply to the pump and disconnect the
power cable from the wall before replacing the air filter.
2. Remove air filter cover, remove filter, and replace with new
filter.
3. Replace air filter cover.
4. Reconnect to wall plug.
7. CLEANING AND DISINFECTION
1. Switch off the electrical supply to the pump and disconnect the power cable from the wall
before cleaning and disinfection.
2. Remove the bedding.
3. If necessary, inflate the mattress.
4. Ensure the underside of the mattress is clear of all sharp objects.
5. Examine the surface of the pump and mattress assembly components for visible blood or body
fluids.
6. Perform one of the followings:
a.
If blood is present, decontaminate the whole mattress product in line with current facility
care protocols.
b.
If blood is not present, remove any soil from the cover with paper towel.
Info: If grossly soiled, the cover should be removed, cleaned and decontaminated.
7. Using a clean sponge or paper towel, wipe down the surface and cells with a diluted detergent
solution or cleaner disinfectant or other germicidal detergent solution.
Please noted do not use
phenol-based cleaning solution.
8. Cleaning and disinfection may be carried out on the cover by hand with hot water and a neutral
detergent or with a sodium hypochlorite (bleach) solution.
9. Alternatively remove the cover and launder at 40°C (104°F), using normal detergent. It is
essential that articles be thoroughly dried after all cleaning procedure and before storage.
10. Perform the following steps to clean the power unit and hose fittings:
a.
Wipe all controls, chassis and hose fittings with a damp cloth and a mild detergent.
b.
Using a nylon brush, gently clean all crevices as they can harbor microorganisms.
c.
Air dry all treated surfaces.
8. TROUBLESHOOTING
The following table can help you determine what may be the causes and solutions to the problem
that you have with your selected Alternating Pressure Mattress System that caused it not to function
as planned. If a problem occurs which is not listed below, contact MOXI for further information.
Air Filter Cover