MAINTENANCE/TROUBLESHOOTING
34
Error Codes
The controller displays an error message (the letter E followed by three digits) if it detects a problem in the system.
If an error occurs during operation, the treadmill decelerates and the walking belt stops. Only the E201 error code
permits the treadmill to keep running. The error code remains on the display until the power is turned off. If an error
code appears, record the code number, then recycle the power. If the error persists, call Mortara Technical Support.
Self-test Errors
Code
Explanation
E101
Controller CPU chip failure
E102
Controller CPU EPROM failure
E103
Controller CPU stuck interrupt
E105
Controller CPU NVRAM failure
Operation Codes
Code
Explanation
E201
Grade system error
E202
Speed check error
E203
Motor is drawing more than the specified current or fan is stopped
E204
The two microcontrollers are not communicating
EPHI
Motor supply voltage too high
EPLO
Motor supply voltage too low