MAINTENANCE/TROUBLESHOOTING
32
Cumulative Usage
The cumulative usage feature lets you determine how often the treadmill is used and how much wear is on the belt
and motor. This can be useful for scheduling service.
Distance
To determine the total distance on each treadmill, follow these steps:
1.
Press
Power
to turn on the treadmill.
2.
Simultaneously press
Stop
and
Slower
on the display panel.
3.
Multiply the number that appears in the multifunctional display by 10 to obtain the total distance in miles.
4.
Record the mileage and the date for your records.
5.
Press
Clear
to reset the display for operation.
Time
To determine the total time of treadmill use, follow these steps:
1.
Press
Power
to turn on the treadmill.
2.
Simultaneously press
Stop
and
Faster
on the display panel.
3.
Multiply the number that appears in the multifunctional display by 10 to obtain the total hours of operation.
4.
Record the total time and the date for your records.
5.
Press
Clear
to reset the display for operation.
Electrical Testing
Electrical testing is to be done by the facility’s biomedical department as required. Check the leakage current of the
treadmill periodically—at least every nine months—to be sure it does not exceed local or provincial standards.